A new positron lifetime spectrometer using a fast digital oscilloscope and BaF2 scintillators

被引:125
作者
Saito, H
Nagashima, Y
Kurihara, T
Hyodo, T
机构
[1] Univ Tokyo, Grad Sch Arts & Sci, Inst Phys, Meguro Ku, Tokyo 1538902, Japan
[2] KEK, High Energy Accelerator Res Org, Inst Mat Struct Sci, Tsukuba, Ibaraki 3050801, Japan
关键词
digital; positron annihilation; gamma-ray; lifetime spectrometer; scintillation detector; digital oscilloscope;
D O I
10.1016/S0168-9002(01)02172-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new positron lifetime spectrometer has been constructed and tested. It consists of a state-of-the-art digital oscilloscope (LeCroy Wavepro 960) and scintillation detectors based on Hamamatsu H3378 (R2083Q) photomultiplier tubes and BaF2 scintillators. A very good time resolution is obtained with scintillators of commonly used sizes (28 mm in diameter, 10-20 mm thick) and wide energy window settings (40-50%). The time resolution of 144 ps full-width at half-maximum (FWHM) is achieved for the positron lifetime measurement, in which time differences are measured between the arrivals of a 1.27 MeV gamma-ray from Na-22 and one of the two positron annihilation gamma-rays of 0.511 MeV. The time resolution is further improved with a 'double-stop' setup, where timing information carried by the positron annihilation gamma-ray pair is combined to determine the time of the annihilation. The time resolution with this setup is 119 ps FWHM. This improvement is interpreted in terms of the energy deposit in the scintillation detectors by gamma-rays. Applications of the new spectrometer to other scintillation timing measurements are also described. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:612 / 617
页数:6
相关论文
共 23 条
[1]  
BAUER W, 1986, POS ANN P 7 INT C PO, P804
[2]  
BECVAR F, 1995, MATER SCI FORUM, V175-, P947, DOI 10.4028/www.scientific.net/MSF.175-178.947
[3]  
Becvár F, 2000, NUCL INSTRUM METH A, V443, P557, DOI 10.1016/S0168-9002(99)01156-0
[4]   CONSTANT FRACTION DIFFERENTIAL DISCRIMINATOR FOR USE IN FAST TIMING COINCIDENCE SYSTEMS [J].
BEDWELL, MO ;
PAULUS, TJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (01) :422-427
[5]  
BEDWELL MO, 1979, POSITRON ANNIHILATIO, P375
[6]  
Bretagnon T., 1992, Materials Science Forum, V105-110, P1841, DOI 10.4028/www.scientific.net/MSF.105-110.1841
[7]   POSITRON LIFETIMES AND LINESHAPE FACTORS IN DEFORMED COPPER [J].
DAUWE, C ;
DORIKENS, M ;
DORIKENS.L ;
SEGERS, D .
APPLIED PHYSICS, 1974, 5 (02) :117-120
[8]   FAST TIMING WITH HAMAMATSU R2083Q PHOTOMULTIPLIERS [J].
DEVRIES, J ;
ZECCA, A ;
BRUSA, RS ;
GRISENTI, RG ;
OSS, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 275 (01) :194-196
[9]  
DEVRIES J, 1989, POS ANN P 8 INT C PO, P654
[10]  
Firestone R. B., 1996, Table of Isotopes, V8th