Lesion contrast enhancement in medical ultrasound imaging

被引:53
作者
Stetson, PF
Sommer, FG
Macovski, A
机构
[1] STANFORD UNIV,DEPT ELECT ENGN,STANFORD,CA 94305
[2] STANFORD UNIV,MED CTR,DEPT DIAGNOST RADIOL,STANFORD,CA 94305
关键词
frequency dependence of ultrasonic backscatter; frequency-diversity ultrasonic imaging; grayscale mapping;
D O I
10.1109/42.611351
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Methods for improving the contrast-to-noise ratio (CNR) of low-contrast lesions in medical ultrasound imaging are described, Differences in the frequency spectra and amplitude distributions of the lesion and its surroundings can be used to increase the CNR of the lesion relative to the background, Automated graylevel mapping is used in combination with a contrast-weighted form of frequency-diversity speckle reduction, in clinical studies, the techniques have yielded mean CNR improvements of 3.2 dB above ordinary frequency-diversity imaging and 5.6 dB over sharper conventional images, with no postprocessing graylevel mapping.
引用
收藏
页码:416 / 425
页数:10
相关论文
共 11 条
[1]  
EDWARDS CH, 1973, ADV CALCULUS SERVERA
[2]  
GEHLBACH SM, 1983, THESIS STANFORD U ST
[3]  
George N., 1972, Optics Communications, V6, P253, DOI 10.1016/0030-4018(72)90187-3
[4]  
Horne R. N., 1990, MATRIX ANAL, P183
[5]  
Ishimaru A., 1996, Wave Propagation and Scattering in Random Media
[6]  
Kino G. S., 1987, ACOUSTIC WAVES DEVIC, P300
[7]   FREQUENCY COMPOUNDING FOR SPECKLE CONTRAST REDUCTION IN PHASED-ARRAY IMAGES [J].
MAGNIN, PA ;
VONRAMM, OT ;
THURSTONE, FL .
ULTRASONIC IMAGING, 1982, 4 (03) :267-281
[8]  
PAPOULIS A, 1965, PROBABILITY RANDOM V, P298
[9]   SPECKLE REDUCTION IN SYNTHETIC-APERTURE RADARS [J].
PORCELLO, LJ ;
MASSEY, NG ;
INNES, RB ;
MARKS, JM .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1976, 66 (11) :1305-1311
[10]  
ROSENFELD A, 1976, DIGITAL PICTURE PROC, P169