Flow properties of liquid calcium alginate polymer injected through medical microcatheters for endovascular embolization

被引:52
作者
Becker, TA
Kipke, DR
机构
[1] Univ Michigan, Dept Biomed Engn, Ann Arbor, MI 48109 USA
[2] Arizona State Univ, Dept Bioengn, Tempe, AZ USA
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 2002年 / 61卷 / 04期
关键词
polymer occlusion; tube flow; shear-thinning; non-Newtonian flow; arteriovenous malformations; aneurysms;
D O I
10.1002/jbm.10202
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The flow properties of liquid calcium alginate injections were investigated for application in endovascular embolization. Alginate shear properties were assessed with a rheometer and a controlled injection system. The experimental results were used to model the flow properties and predict alginate's flow characteristics within various medical microcatheter delivery systems, The results suggest that alginates undergo shear-thinning effects with increasing shear. A flow comparison of 2.0 wt % alginate and a Newtonian fluid (82 cP) injected from the same microcatheter had similar flow rates at low injection pressure (100 kPa). However, at high injection pressure (2100 kPa), the alginate was injectable at a flow rate 100% higher than was the Newtonian fluid. Further analysis of injections through microcatheters resulted in a flow model for predicting viscosity changes, flow rates, and injection pressures of liquid alginate at medium-to-high shear rates. The predicted injection pressures and flow rates had an average variance of less than 15% from that of the experimental flow data. This study indicates that calcium alginate has the requisite flow properties for successful delivery to vascular lesions via endovascular injection. Possible uses of alginates include treating arteriovenous malformations (AVMs), aneurysms, blood flow to tumors, and vascular hemorrhages. (C) 2002 Wiley Periodicals, Inc.
引用
收藏
页码:533 / 540
页数:8
相关论文
共 30 条
[1]  
Becker TA, 2001, J BIOMED MATER RES, V54, P76, DOI 10.1002/1097-4636(200101)54:1<76::AID-JBM9>3.0.CO
[2]  
2-V
[3]  
CAHILL M, 1997, PHYSICIANS DRUG HDB
[4]  
Casasco A, 1999, AM J NEURORADIOL, V20, P179
[5]  
Dawson RC, 1998, AM J NEURORADIOL, V19, P571
[6]  
Deveikis JP, 1998, NEUROIMAG CLIN N AM, V8, P401
[7]  
DeVos P, 1997, DIABETOLOGIA, V40, P262
[8]   Effect of the alginate composition on the biocompatibility of alginate-polylysine microcapsules [J].
DeVos, P ;
DeHaan, B ;
VanSchilfgaarde, R .
BIOMATERIALS, 1997, 18 (03) :273-278
[9]  
DORNISH M, 1997, KEY CHARACTERIZATION
[10]   Relationship between viscosity and determined injection pressure in angiography catheters for common roentgen contrast media [J].
Dyvik, K ;
Dyrstad, K ;
Tronstad, A .
ACTA RADIOLOGICA, 1995, 36 :43-49