Pending removal from the market of a commercial assay (the AUSAB [Abbott Laboratories] enzyme immunoassay [EIA]) for the determination of antibodies to hepatitis B surface antigen (HBsAg), a new in-house quantitative enzyme-linked immunosorbent assay (ELISA) to measure antibodies against HBsAg (anti-HBs) was developed (anti-HBs in-house). Specific anti-HBs antibodies were sandwiched between the precoated HBsAg ad and ay subtypes purified from plasma from hepatitis B virus (HBV) human carriers and the recombinant HBsAg adw2 subtype tagged with horseradish peroxidase. The assay was calibrated against the 1st International Reference Preparation for anti-hepatitis B immunoglobulin (lot 1977-W1042). Analytical sensitivity and the limit of quantitation were estimated at 0.43 mIU/ml and 2.0 mIU/ml, respectively. Overall reproducibility was 11.86%, and accuracy was estimated to be 94.89%. More than 4,000 samples from seven clinical trials were tested with the anti-HBs in-house assay and compared to results generated with AUSAB EIA and AUSAB radioimmunoassay (RIA). During the technical validation, the anti-HBs in-house assay was compared to the AUSAB RIA as a reference (n = 919). Overall assessment of concordance and Deming's regression analysis were performed. The coefficient of correlation between the AUSAB RIA and anti-HBs in-house assay was 0.9815 with a slope of 0.9187. The overall agreement between anti-HBs in-house and AUSAB RIA was 97.61%, considering the clinical cutoffs at 3.3 mIU/ml and 1.0 mIU/ml for the respective assays. From a clinical perspective, seroprotection rates and anti-HBs geometric mean antibody concentrations for individual studies calculated with either the in-house assay or the reference assays were similar. Conclusions of individual studies were confirmed. The performance characteristics of the in-house assay are acceptable. There is no evidence that use of the new assay would lead to different clinical conclusions from the reference method.
引用
收藏
页码:1236 / 1246
页数:11
相关论文
共 16 条
[1]
[Anonymous], 2004, WEEKLY EPIDEMIOLOGIC, V28, P255
[2]
[Anonymous], 2001, Guidance for industry, bioanalytical method validation
[3]
Barker L, 1977, EXPERT COMMITTEE BIO
[4]
A strategy for validation of bioanalytical methods [J].
Braggio, S ;
Barnaby, RJ ;
Grossi, P ;
Cugola, M .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1996, 14 (04) :375-388
[5]
Validation of immunoassays for bioanalysis: a pharmaceutical industry perspective [J].
Findlay, JWA ;
Smith, WC ;
Lee, JW ;
Nordblom, GD ;
Das, I ;
DeSilva, BS ;
Khan, MN ;
Bowsher, RR .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2000, 21 (06) :1249-1273
[6]
HAUSER P, 1987, POSTGRAD MED J, V63, P83
[7]
Recombinant hepatitis B vaccine (Engerix-B®) -: A review of its immunogenicity and protective efficacy against hepatitis B [J].
Keating, GM ;
Noble, S .
DRUGS, 2003, 63 (10) :1021-1051
[8]
Workshop on bioanalytical methods validation for macromolecules: Summary report [J].
Miller, KJ ;
Bowsher, RR ;
Celniker, A ;
Gibbons, J ;
Gupta, S ;
Lee, JW ;
Swanson, SJ ;
Smith, WC ;
Weiner, RS .
PHARMACEUTICAL RESEARCH, 2001, 18 (09) :1373-1383
[9]
PEROXIDASE-LABELED ANTIBODY - NEW METHOD OF CONJUGATION [J].
NAKANE, PK ;
KAWAOI, A .
JOURNAL OF HISTOCHEMISTRY & CYTOCHEMISTRY, 1974, 22 (12) :1084-1091
[10]
PETRE J, 1987, POSTGRAD MED J, V63, P73