Determination of copper, lead and cadmium in whole blood by stripping voltammetry with the use of graphite electrodes

被引:27
作者
Brainina, K [1 ]
Schafer, H [1 ]
Ivanova, A [1 ]
Khanina, R [1 ]
机构
[1] METROHM LTD,CH-9101 HERISAU,SWITZERLAND
关键词
copper; lead; cadmium; blood; voltammetry;
D O I
10.1016/0003-2670(96)00181-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The problem with toxic metal ion determination in blood is the adsorption of organic compounds on the electrode surface and the formation of complexes between metal ions and organic constituents of blood. This is the reason why usually preliminary acid digestion or other sample pretreatment is used. Two kinds of electrodes have been used: ''Ultra-Trace Electrode'', made from impregnated graphite (I), and thick film graphite disposable electrodes (II). The analysis of whole blood with different sample preparation methods shows, that chemical digestion is not necessary for the analysis. Electrochemical two-stage sample preparation provides the possibility for analysing whole blood with the mentioned electrodes. Thick film disposable electrodes are less sensitive to the interference of organic constituents of blood. These electrodes give the possibility to determine total cadmium, lead and copper concentration in whole blood without special sample pretreatment. The application of ''Ultra-Trace Electrode'' for blood analysis is possible only after preliminary pretreatment of blood by chemical digestion or electrochemical sample preparation.
引用
收藏
页码:175 / 181
页数:7
相关论文
共 15 条
  • [1] AUTOMATED-DETERMINATION OF CADMIUM AND LEAD IN WHOLE-BLOOD BY COMPUTERIZED FLOW POTENTIOMETRIC STRIPPING WITH CARBON-FIBER ELECTRODES
    ALMESTRAND, L
    JAGNER, D
    RENMAN, L
    [J]. ANALYTICA CHIMICA ACTA, 1987, 193 : 71 - 79
  • [2] [Anonymous], TECHNOGENESIS PYRITE
  • [3] BOECKX RL, 1986, ANAL CHEM, V58, P275
  • [4] BRAININA K, 1993, ELECTROANALYTICAL ST
  • [5] CHARACTERIZATION OF THICK-LAYER GRAPHITE DISPOSABLE VOLTAMMETRIC ELECTRODES
    BRAININA, KZ
    BOND, AM
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (15) : 2586 - 2591
  • [6] BRAININA KZ, 1985, ANAL LETT PT A, V18, P117
  • [7] BRAININA KZ, 1988, STRIPPING ELECTROANA, P114
  • [8] *ENV SCI ASS, 1975, TMA1 ENV SCI ASS
  • [9] ERSHOV YA, 1989, MECH TOXIC EFFECTS I
  • [10] DETERMINATION OF LEAD IN BLOOD BY SQUARE-WAVE ANODIC-STRIPPING VOLTAMMETRY AT A CARBON DISK ULTRAMICROELECTRODE
    FELDMAN, BJ
    OSTERLOH, JD
    HATA, BH
    DALESSANDRO, A
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (13) : 1983 - 1987