Intentional Leaks in Industrial Masks Have a Significant Impact on Efficacy of Bilevel Noninvasive Ventilation A Bench Test Study

被引:55
作者
Borel, Jean Christian [2 ]
Sabil, AbdelKebir [3 ]
Janssens, Jean-Paul [4 ]
Couteau, Marie [5 ]
Boulon, Loren [3 ]
Levy, Patrick [2 ]
Pepin, Jean-Louis [1 ,2 ]
机构
[1] CHU Grenoble, Lab EFCR, F-38043 Grenoble 09, France
[2] Univ Grenoble 1, INSERM, ERI 0017, Lab HP2, Grenoble, France
[3] Assoc Med Tech AGIR Dom, Meylan, France
[4] Hop Cantonal Univ Geneva, Serv Pneumol, Geneva, Switzerland
[5] Grp COVIDIEN, Soc AIROX, Pau, France
关键词
bench test study; bilevel pressure cycled ventilators; masks for noninvasive ventilation; performance of ventilators; PRESSURE SUPPORT VENTILATION; MECHANICAL VENTILATION; PERFORMANCE-CHARACTERISTICS; OBESITY-HYPOVENTILATION; LUNG MODEL; SLEEP; HOME; ASYNCHRONY; PATTERNS; ASSISTANCE;
D O I
10.1378/chest.08-1340
中图分类号
R4 [临床医学];
学科分类号
100218 [急诊医学];
摘要
Background: During noninvasive ventilation, nonintentional leaks have a detrimental effect on the efficacy of ventilation. A wide range of industrial masks are available, with intentional leaks of different importance. The potential impact of this variability in intentional leaks on performances of bilevel ventilators has not been assessed. Objective: To measure intentional leaks in seven different industrial masks and determine whether higher leaks modify ventilator performance and quality of ventilation. Methods: Seven interfaces connected to four ventilators, the VPAP III. ST (ResMed; NorthRyde, Australia), the BiPAP Harmony (Respironics; Monroeville, PA), the SmartAir ST (Covidien/Airox; Pau, France), and the GoodKnight 425 ST Bilevel (Covidien/Tyco-Nellcor/Puritan Bennett; Pleasanton, CA), were adapted on a mannequin connected to a lung model (ASL5000, IngMar Medical; Pittsburgh, PA). Inspiratory positive airway pressure (IPAP) and expiratory positive airway pressure were 14 and 4 cm H2O, respectively. The lung model was set with a respiratory rate of 15 cycles per min and a duration of inspiration of 1 s in three simulated conditions (normal, restrictive, and obstructive). Inspiratory trigger delay and effort, capacity to achieve and maintain IPAP, expiratory cycling and tidal volume were analyzed for all masks and ventilators in the three simulated lung conditions. Results: The level of intentional leaks in the seven masks ranged from 30 to 45 L/min for an IPAP of 14 cm H2O. Importance of leaks did not influence trigger performances. However, capacity to achieve and maintain IPAP was significantly decreased with all ventilators and in all simulated lung conditions when intentional leaks increased. This led to a maximum reduction in delivered tidal volume of 48 mL. Expiratory cycling was not affected by the level of intentional leaks except in obstructive lung conditions. Conclusion: Mask intentional leaks can impair efficacy of ventilation, especially when > 40 L/min.
引用
收藏
页码:669 / 677
页数:9
相关论文
共 25 条
[1]
Performance characteristics of 10 home mechanical ventilators in pressure-support mode - A comparative bench study [J].
Battisti, A ;
Tassaux, D ;
Janssens, JP ;
Michotte, JB ;
Jaber, S ;
Jolliet, P .
CHEST, 2005, 127 (05) :1784-1792
[2]
Performance characteristics of bilevel pressure ventilators - A lung model study [J].
Bunburaphong, T ;
Imanaka, H ;
Nishimura, M ;
Hess, D ;
Kacmarek, RM .
CHEST, 1997, 111 (04) :1050-1060
[3]
Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger [J].
Calderini, E ;
Confalonieri, M ;
Puccio, PG ;
Francavilla, N ;
Stella, L ;
Gregoretti, C .
INTENSIVE CARE MEDICINE, 1999, 25 (07) :662-667
[4]
Comparative evaluation of different helmets on patient-ventilator interaction during noninvasive ventilation [J].
Costa, R. ;
Navalesi, P. ;
Spinazzola, G. ;
Rossi, M. ;
Cavaliere, F. ;
Antonelli, M. ;
Proietti, R. ;
Conti, G. .
INTENSIVE CARE MEDICINE, 2008, 34 (06) :1102-1108
[5]
The interface: crucial for successful noninvasive ventilation [J].
Elliott, MW .
EUROPEAN RESPIRATORY JOURNAL, 2004, 23 (01) :7-8
[6]
Effect of sleep on patient/ventilator asynchrony in patients undergoing chronic non-invasive mechanical ventilation [J].
Fanfulla, Francesco ;
Taurino, Anna Eugenia ;
Lupo, Nadia D'Artavilla ;
Trentin, Rossella ;
D'Ambrosio, Carolyn ;
Nava, Stefano .
RESPIRATORY MEDICINE, 2007, 101 (08) :1702-1707
[7]
FERGUSON GT, 1995, AM J RESP CRIT CARE, V151, P1126
[8]
Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study [J].
Ferreira, Juliana C. ;
Chipman, Daniel W. ;
Kacmarek, Robert M. .
INTENSIVE CARE MEDICINE, 2008, 34 (09) :1669-1675
[9]
Air leaks during mechanical ventilation as a cause of persistent hypercapnia in neuromuscular disorders [J].
Gonzalez, J ;
Sharshar, T ;
Hart, N ;
Chadda, K ;
Raphaël, JC ;
Lofaso, F .
INTENSIVE CARE MEDICINE, 2003, 29 (04) :596-602
[10]
Respiratory patterns during sleep in obesity-hypoventilation patients treated with nocturnal pressure support - A preliminary report [J].
Guo, Yan Fei ;
Sforza, Emilia ;
Janssens, Jean Paul .
CHEST, 2007, 131 (04) :1090-1099