Grinding precision and accuracy of fit of CEREC 2 CAD-CIM inlays

被引:47
作者
Mormann, WH
Schug, J
机构
[1] Div. Aesthetic Comp. Restorations, Clin. Prev. Dent., Periodontol. G., University of Zurich, CH-8028 Zurich
[2] Found. for Comp.-Assisted Dentistry, Zürich
关键词
D O I
10.14219/jada.archive.1997.0025
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
The authors conducted research to determine the grinding precision and accuracy of fit of ceramic inlays generated with the completely redesigned CEREC 2 computer-assisted design/computer-integrated manufacturing, or CAD-CIM, unit. They found that the grinding precision of the CEREC 2 unit was 2.4 times greater than that of CEREC 1. With CEREC 2, the mean (+/- standard deviation) luting interfaces were 56 +/- 27 micrometers, which is a 30 percent improvement in the accuracy of fit compared with that of CEREC 1.
引用
收藏
页码:47 / 53
页数:7
相关论文
共 41 条
[1]
Benz C, 1996, CAD CIM AESTHETIC DE, P516
[2]
Besek MJ, 1996, CAD/CIM IN AESTHETIC DENTISTRY, P347
[3]
Brauner AW, 1996, CAD/CIM IN AESTHETIC DENTISTRY, P217
[4]
Calamia J R, 1994, Curr Opin Cosmet Dent, P67
[5]
Duret F, 1993, J Can Dent Assoc, V59, P445
[6]
Feher A, 1991, Schweiz Monatsschr Zahnmed, V101, P857
[7]
FETT HP, 1991, QUINTESSENCE INT, V22, P823
[8]
FURRER O, 1993, SCHWEIZ MONATSSCHR Z, V103, P17
[9]
MARGINAL ACCURACY OF CAD/CAM INLAYS MADE WITH THE ORIGINAL AND THE UPDATED SOFTWARE [J].
INOKOSHI, S ;
VANMEERBEEK, B ;
WILLEMS, G ;
LAMBRECHTS, P ;
BRAEM, M ;
VANHERLE, G .
JOURNAL OF DENTISTRY, 1992, 20 (03) :171-177
[10]
Isenberg B P, 1992, J Esthet Dent, V4, P173, DOI 10.1111/j.1708-8240.1992.tb00690.x