On-line separation and preconcentration of chromium species in seawater

被引:90
作者
Posta, J
Alimonti, A
Petrucci, F
Caroli, S
机构
[1] IST SUPER SANITA,I-00161 ROME,ITALY
[2] LAJOS KOSSUTH UNIV,DEPT INORGAN & ANALYT CHEM,H-4010 DEBRECEN 10,HUNGARY
关键词
inductively coupled plasma mass spectrometry; inductively coupled plasma atomic emission spectrometry; ultrasonic nebulization; chromium speciation; chromium(VI) preconcentration; seawater;
D O I
10.1016/0003-2670(96)00030-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A study was undertaken to ascertain the analytical capabilities of combined HPLC-ICP-AES with ultrasonic nebulization or HPLC-ICP-MS systems to perform on-line separation of Cr-III-Cr-VI and on-line preconcentration of Cr-VI Tetrabutylammonium bromide was used as the ion-pair forming agent and a reverse phase C-18 column was used for separation. By optimizing the conditions of separation and preconcentration as well as plasma parameters a method could be developed for the determination of Cr-III and Cr-VI in seawater by HPLC-ICP-AES with ultrasonic nebulization. In this matrix detection limits (3 sigma criterion) turned out to be 4.6 ng/ml for Cr-VI and 3.7 ng/ml for Cr-VI, while the RSD is 2%-3% at the level of 100 ng/ml Ct. As the preconcentration of Cr-VI is regarded, the detection limit of ICP-AES depends, among others, on the sample volume loaded onto the column. It resulted to be 0.20 ng/ml for an initial volume of 2 ml, while the RSD at 10 ng/ml Cr-VI was +/- 3.8%. On the other hand, the high salinity of seawater and C content of the separation eluent pose serious problems with ICP-MS detection, In fact, in this matrix only Cr-VI preconcentration was achieved with a detection limit of 0.12 ng/ml for Cr-VI (initial volume 2 ml). Due to isobaric interferences, the isotope at mass 50 (relative abundance 4.31%) was preferred.
引用
收藏
页码:185 / 193
页数:9
相关论文
共 13 条
[1]  
Bowen H. M., 1979, ENV CHEM ELEMENTS
[2]   PRECONCENTRATION AND DETERMINATION OF TRACE CHROMIUM(III) BY FLOW-INJECTION INDUCTIVELY-COUPLED PLASMA/ATOMIC EMISSION-SPECTROMETRY [J].
COX, AG ;
MCLEOD, CW .
ANALYTICA CHIMICA ACTA, 1986, 179 :487-490
[3]   RAPID SEQUENTIAL DETERMINATION OF CHROMIUM(III) - CHROMIUM(VI) BY FLOW-INJECTION ANALYSIS - INDUCTIVELY COUPLED PLASMA ATOMIC-EMISSION SPECTROMETRY [J].
COX, AG ;
COOK, IG ;
MCLEOD, CW .
ANALYST, 1985, 110 (04) :331-333
[4]   TRACE ANALYSIS AND SPECIATION FOR CR(VI) AND CR(III) VIA HPLC-DIRECT CURRENT PLASMA EMISSION-SPECTROSCOPY (HPLC-DCP) [J].
KRULL, IS ;
PANARO, KW ;
GERSHMAN, LL .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1983, 21 (10) :460-472
[5]   SPECIATION OF CR (III) AND CR (VI) VIA REVERSED PHASE HPLC WITH INDUCTIVELY COUPLED PLASMA EMISSION SPECTROSCOPIC DETECTION (HPLC-ICP) [J].
KRULL, IS ;
BUSHEE, D ;
SAVAGE, RN ;
SCHLEICHER, RG ;
SMITH, SB .
ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1982, 15 (03) :267-281
[6]  
MILOSAVLJEVIC EB, 1985, MIKROCHIM ACTA, V3, P353
[7]  
Nriagu J.O., 1988, Chromium in the natural and human environments
[8]   HIGH-PERFORMANCE FLOW FLAME ATOMIC-ABSORPTION SPECTROMETRY FOR AUTOMATED ONLINE SEPARATION AND DETERMINATION OF CR(III) CR(VI) AND PRECONCENTRATION OF CR(VI) [J].
POSTA, J ;
BERNDT, H ;
LUO, SK ;
SCHALDACH, G .
ANALYTICAL CHEMISTRY, 1993, 65 (19) :2590-2595
[9]   DETERMINATION OF CHROMIUM BY ONLINE PRECONCENTRATION ON A POLY(HYDROXAMIC ACID) RESIN IN FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY [J].
SHAH, A ;
DEVI, S .
ANALYTICA CHIMICA ACTA, 1990, 236 (02) :469-473
[10]   DIFFERENTIAL DETERMINATION OF CHROMIUM(VI) AND TOTAL CHROMIUM IN NATURAL-WATERS USING FLOW-INJECTION ONLINE SEPARATION AND PRECONCENTRATION ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
SPERLING, M ;
YIN, XF ;
WELZ, B .
ANALYST, 1992, 117 (03) :629-635