The intubating laryngeal mask .1. Development of a new device for intubation of the trachea

被引:183
作者
Brain, AIJ
Verghese, C
Addy, EV
Kapila, A
机构
[1] UNIV LONDON, INST LARYNGOL, LONDON, ENGLAND
[2] OXFORD RADCLIFFE NHS TRUST, OXFORD, ENGLAND
关键词
equipment; masks anaesthesia; intubation tracheal; technique;
D O I
10.1093/bja/79.6.699
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The standard laryngeal mask airway (LMA) functions both as a ventilatory device and as an aid to blind/fibrescopic-guided tracheal intubation. We describe the radiological and laboratory work used to bioengineer a new laryngeal mask prototype, the intubating laryngeal mask airway (ILMA). The aim was to create a new airway system with better intubation characteristics than the LMA. Other design goals were to eliminate the need for head-neck manipulation and insertion of fingers in the mouth during placement. Development was aided by analysis of magnetic resonance images of the human pharynx and laboratory testing with a variety of tracheal tubes. The principal features of this new system are an anatomically curved, rigid airway tube with an integral guiding handle, an epiglottic elevating bar replacing the mask bars, a guiding ramp built into the floor of the mask aperture and a modified silicone tracheal tube developed for use with the device.
引用
收藏
页码:699 / 703
页数:5
相关论文
共 13 条
[1]   THE DISTANCE BETWEEN THE GRILLE OF THE LARYNGEAL MASK AIRWAY AND THE VOCAL CORDS - IS CONVENTIONAL INTUBATION THROUGH THE LARYNGEAL MASK SAFE [J].
ASAI, T ;
LATTO, IP ;
VAUGHAN, RS .
ANAESTHESIA, 1993, 48 (08) :667-669
[2]   Laryngeal mask airway and the ASA difficult airway algorithm [J].
Benumof, JL .
ANESTHESIOLOGY, 1996, 84 (03) :686-699
[3]  
BENUMOF JL, 1991, ANESTHESIOLOGY, V75, P1087
[4]   The intubating laryngeal mask .2. a preliminary clinical report of a new means of intubating the trachea [J].
Brain, AIJ ;
Verghese, C ;
Addy, EV ;
Kapila, A ;
Brimacombe, J .
BRITISH JOURNAL OF ANAESTHESIA, 1997, 79 (06) :704-709
[5]   THE PRESSOR-RESPONSE AND LARYNGEAL MASK INSERTION - A COMPARISON WITH TRACHEAL INTUBATION [J].
BRAUDE, N ;
CLEMENTS, EAF ;
HODGES, UM ;
ANDREWS, BP .
ANAESTHESIA, 1989, 44 (07) :551-554
[6]   THE SPLIT LARYNGEAL MASK AIRWAY [J].
BRIMACOMBE, J .
ANAESTHESIA, 1993, 48 (07) :639-639
[7]   ADVERSE RESPIRATORY EVENTS IN ANESTHESIA - A CLOSED CLAIMS ANALYSIS [J].
CAPLAN, RA ;
POSNER, KL ;
WARD, RJ ;
CHENEY, FW .
ANESTHESIOLOGY, 1990, 72 (05) :828-833
[8]  
HEATH ML, 1991, EUR J ANAESTH, P41
[9]   INTUBATION LESIONS OF LARYNX [J].
KAMBIC, V ;
RADSEL, Z .
BRITISH JOURNAL OF ANAESTHESIA, 1978, 50 (06) :587-590
[10]  
KAPILA A, 1995, BRIT J ANAESTH, V75, pP228