Video Laryngoscopy with the Macintosh Video Laryngoscope in Simulated Prehospital Scenarios by Paramedic Students

被引:33
作者
Aziz, Michael [1 ]
Dillman, Dawn [1 ]
Kirsch, Jeffrey R. [1 ]
Brambrink, Ansgar [1 ]
机构
[1] Oregon Hlth & Sci Univ, Dept Anesthesiol & Perioperat Med, Portland, OR 97239 USA
关键词
video laryngoscopy; simulation; paramedic; Macintosh video laryngoscope; prehospital; intubation; TRACHEAL INTUBATION; GLIDESCOPE(R); VIDEOLARYNGOSCOPE;
D O I
10.1080/10903120802706070
中图分类号
R4 [临床医学];
学科分类号
100218 [急诊医学];
摘要
Objective. To investigate the effectiveness of the Karl Storz BERCI DCI Macintosh video laryngoscope (MVL) via the TELE PACK system for facilitating intubation by novice paramedic students in a simulation environment. We assessed the laryngeal view, measured by percentage of glottic opening (POGO), when intubating the SimMan manikin airway in different settings. The primary endpoint was the best POGO achieved by the student. Secondary endpoints included intubation times and success rate. Method. We enrolled 25 novice paramedic students to intubate SimMan manikins. Students were randomized to use either a conventional Macintosh 3 (Mac3) blade alone or the MVL with a Mac3 blade. Students attempted their first intubation with the manikin on a stretcher in a normal neck position and reattempted intubation with the manikin's neck stiffened. The groups then crossed over using the alternate device to repeat the attempts in the manikin with a normal neck and with a stiffened neck. The students then attempted the same sequence of four intubations with the manikin on the floor. Results. The MVL significantly improved POGO in all scenarios (p 0.05). The MVL improved mean POGO 16% 6% in the manikin with a normal neck position on a stretcher and 33% 7% in the manikin with a stiff neck on the floor. The improvement was significantly greater in simulated difficult scenarios. The intubation success rate (94%) was equal in the two groups, and the POGO was significantly worse in the failures. In some subgroups, intubation times were longer with the MVL. Conclusion. The MVL improves the laryngeal view for novice laryngoscopists in a simulated setting, and this improvement is greatest in simulated difficult scenarios.
引用
收藏
页码:251 / 255
页数:5
相关论文
共 16 条
[1]
Optimization of glottic exposure during intubation of a patient lying supine on the ground [J].
Adnet, F ;
Lapostolle, F ;
Borron, SW ;
Hennequin, B ;
Leclercq, F ;
Fleury, M .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1997, 15 (06) :555-557
[2]
Cooper RM, 2005, CAN J ANAESTH, V52, P191, DOI 10.1007/BF03027728
[3]
Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study [J].
Enomoto, Y. ;
Asai, T. ;
Arai, T. ;
Kamishima, K. ;
Okuda, Y. .
BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (04) :544-548
[4]
Cervical spine movement during laryngoscopy using the Airway Scope compared with the Macintosh laryngoscope [J].
Hirabayashi, Y. ;
Fujita, A. ;
Seo, N. ;
Sugimoto, H. .
ANAESTHESIA, 2007, 62 (10) :1050-1055
[5]
Comparison of direct and video-assisted views of the larynx during routine intubation [J].
Kaplan, Marshal B. ;
Hagberg, Carin A. ;
Ward, Denham S. ;
Brambrink, Ansgar ;
Chhibber, Ashwani K. ;
Heidegger, Thomas ;
Lozada, Leonardo ;
Ovassapian, Andranik ;
Parsons, David ;
Ramsay, James ;
Wilhelm, Wolfram ;
Zwissler, Bernhard ;
Gerig, Haus J. ;
Hofstetter, Christian ;
Karan, Suzanne ;
Kreisler, Nevin ;
Pousman, Robert M. ;
Thierbach, Andreas ;
Wrobel, Marc ;
Berci, George .
JOURNAL OF CLINICAL ANESTHESIA, 2006, 18 (05) :357-362
[6]
Seeing is believing: the importance of video laryngoscopy in teaching and in managing the difficult airway [J].
Kaplan, MB ;
Ward, D ;
Hagberg, CA ;
Berci, G ;
Hagiike, M .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2006, 20 (Suppl 2) :479-483
[7]
Koetter K P, 1997, Prehosp Emerg Care, V1, P96, DOI 10.1080/10903129708958796
[8]
Assessment of airway visualization: Validation of the percentage of glottic opening (POGO) scale [J].
Levitan, RM ;
Ochroch, EA ;
Kush, S ;
Shofer, FS ;
Hollander, JE .
ACADEMIC EMERGENCY MEDICINE, 1998, 5 (09) :919-923
[9]
Ease of intubation with the GlideScope or Macintosh laryngoscope by inexperienced operators in simulated difficult airways [J].
Lim, Y ;
Lim, TJ ;
Liu, EHC .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2004, 51 (06) :641-642
[10]
The use of the BERCI DCI® Video Laryngoscope for teaching novices direct laryngoscopy and tracheal intubation [J].
Low, D. ;
Healy, D. ;
Rasburn, N. .
ANAESTHESIA, 2008, 63 (02) :195-201