Sources of registration error with image guidance systems during endoscopic anterior cranial base surgery

被引:29
作者
Snyderman, C
Zimmer, LA
Kassam, A
机构
[1] Univ Pittsburgh, Med Ctr, Dept Otolaryngol Head & Neck Surg, Pittsburgh, PA 15213 USA
[2] Univ Pittsburgh, Med Ctr, Dept Neurosurg, Pittsburgh, PA 15213 USA
关键词
D O I
10.1016/j.otohns.2004.03.002
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
OBJECTIVES: The goal of this study was to evaluate the accuracy of the registration process and to identity potential sources of error during anterior cranial base surgery. STUDY DESIGN AND SETTING: The registration accuracy of image guidance and the location of excluded fiducials were recorded prospectively from 50 endoscopic, anterior cranial base procedures in an academic university setting. RESULTS: The mean error of initial registration was 2.8 mm (range, 1.4 to 7.1 mm). Following the exclusion of fiducials the mean error of registration was 1.6 mm (range, 0.6 to 3.7 mm). There was a significant improvement in the mean error rate from initial to final registration following the exclusion of fiducials (P < 0.0001). Posterior fiducials were excluded most often and anterior fiducials were excluded the least. Registration accuracy was similar for CT and MRI (P = 0.64). CONCLUSIONS: The accuracy of the Stryker Image Guidance System is enhanced by the exclusion of individual fiducials with high registration errors. SIGNIFICANCE: The exclusion of fiducials with high registration errors increases the accuracy of image guidance in anterior cranial. base surgery.
引用
收藏
页码:145 / 149
页数:5
相关论文
共 14 条
[1]   Comparison of registration accuracy of skin- and bone-implanted fiducials for frameless stereotaxis of the brain: A prospective study [J].
Ammirati, M ;
Gross, JD ;
Ammirati, G ;
Dugan, S .
SKULL BASE-AN INTERDISCIPLINARY APPROACH, 2002, 12 (03) :125-131
[2]  
ANON JB, 1994, LARYNGOSCOPE, V104, P901
[3]   Frameless stereotaxy with scalp-applied fiducial markers for brain biopsy procedures: experience in 218 cases [J].
Barnett, GH ;
Miller, DW ;
Weisenberger, J .
JOURNAL OF NEUROSURGERY, 1999, 91 (04) :569-576
[4]   Computer-assisted intraoperative navigation during skull base surgery [J].
Carrau, RL ;
Snyderman, CH ;
Curtin, HD ;
Janecka, IP ;
Stechison, M ;
Weissman, JL .
AMERICAN JOURNAL OF OTOLARYNGOLOGY, 1996, 17 (02) :95-101
[5]  
Grunert P, 1998, Comput Aided Surg, V3, P166, DOI 10.1002/(SICI)1097-0150(1998)3:4<166::AID-IGS6>3.3.CO
[6]  
2-5
[7]   REGISTRATION ERROR QUANTIFICATION OF A SURFACE-BASED MULTIMODALITY IMAGE FUSION SYSTEM [J].
HEMLER, PF ;
NAPEL, S ;
SUMANAWEERA, TS ;
PICHUMANI, R ;
VANDENELSEN, PA ;
MARTIN, D ;
DRACE, J ;
ADLER, JR ;
PERKASH, I .
MEDICAL PHYSICS, 1995, 22 (07) :1049-1056
[8]   A passive-marker-based optical system for computer-aided surgery in otorhinolaryngology:: Development and first clinical experiences [J].
Klimek, L ;
Ecke, U ;
Lübben, B ;
Witte, J ;
Mann, W .
LARYNGOSCOPE, 1999, 109 (09) :1509-1515
[9]   A comparison of image guidance systems for sinus surgery [J].
Metson, R ;
Gliklich, RE ;
Cosenza, M .
LARYNGOSCOPE, 1998, 108 (08) :1164-1170
[10]   In vivo accuracy testing and clinical experience with the ISG viewing wand [J].
Sipos, EP ;
Tebo, SA ;
Zinreich, SJ ;
Long, DM ;
Brem, H .
NEUROSURGERY, 1996, 39 (01) :194-202