Hemolytic effect of surface roughness of an impeller in a centrifugal blood pump

被引:15
作者
Takami, Y
Nakazawa, T
Makinouchi, K
Tayama, E
Glueck, J
Benkowski, R
Nose, Y
机构
[1] Department of Surgery, Baylor College of Medicine, Houston, TX
[2] Department of Surgery, Baylor College of Medicine, Houston, TX 77030, One Baylor Plaza
关键词
centrifugal blood pump; impeller; surface roughness; hemolysis; cardiopulmonary bypass;
D O I
10.1111/j.1525-1594.1997.tb03723.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The present study investigates how the surface roughness of an impeller affects hemolysis in the pivot bearing supported Gyro C1E3 pump. This study focuses on particular areas of the impeller surface in the impeller type centrifugal pump. Seven Gyro C1E3 pumps were prepared with smooth surface housings and different impeller parts with different surface roughnesses. The vanes, top side, and backside of the impeller were independently subjected to vapor polishing, fine sand blasting, or coarse sand blasting to produce three different grades of surface roughness. These surfaces were then examined by a surface profile instrument. Using these pumps with different impellers, in vitro hemolysis tests were performed simulating cardiopulmonary bypass (5 L/min, 350 mm Hg). The findings of this study conclusively proved that surface roughness of the back side of the impeller has the greatest effect on hemolysis, followed by the top side and then the vanes. The following are reasons for these findings. First, the shear rate may be greater on the back side than on the top side because of the smaller gap between the back and the housing and the greater relative speed against the impeller. Second, the fluid beneath the impeller may have a longer exposure time because there is little chance for the fluid to mix beneath the impeller. Third, the shear rate may be greater on the top side of the impeller than on the vanes because a vortex formation occurs behind the vanes.
引用
收藏
页码:686 / 690
页数:5
相关论文
共 11 条
[1]  
Bennett J.M., 1989, INTRO SURFACE ROUGHN
[2]  
DEISS A, 1993, WINTROBES CLIN HEMAT
[3]   DEVELOPMENT AND EVALUATION OF ANTITHROMBOGENIC CENTRIFUGAL PUMP - THE BAYLOR-C-GYRO-PUMP-ECCENTRIC-INLET-PORT-MODEL [J].
OHARA, Y ;
MAKINOUCHI, K ;
GLUECK, J ;
SUTHERLAND, B ;
SHIMONO, T ;
NAITO, K ;
TASAI, K ;
ORIME, Y ;
TAKATANI, S ;
NOSE, Y .
ARTIFICIAL ORGANS, 1994, 18 (09) :673-679
[4]  
OHARA Y, 1993, ARTIF ORGANS, V17, P599
[5]   AN ULTIMATE, COMPACT, SEALLESS CENTRIFUGAL VENTRICULAR ASSIST DEVICE - BAYLOR C-GYRO PUMP [J].
OHARA, Y ;
MAKINOUCHI, K ;
ORIME, Y ;
TASAI, K ;
NAITO, K ;
MIZUGUCHI, K ;
SHIMONO, T ;
DAMM, G ;
GLUECK, J ;
TAKATANI, S ;
NOON, GP ;
NOSE, Y .
ARTIFICIAL ORGANS, 1994, 18 (01) :17-24
[6]  
Sakuma I, 1993, ASAIO J, V39, pM433
[7]  
Sakuma Ichiro, 1994, Japanese Journal of Artificial Organs, V23, P30
[8]  
TAKAMI Y, 1996, IN PRESS ARTIF ORGAN, V20
[9]  
TAKAMI Y, IN PRESS ARTIF ORGAN
[10]  
UMEZU M, 1996, IN PRESS ARTIF ORGAN, V20