Frequency-domain techniques enhance optical mammography: Initial clinical results

被引:299
作者
Franceschini, MA
Moesta, KT
Fantini, S
Gaida, G
Gratton, E
Jess, H
Mantulin, WW
Seeber, M
Schlag, PM
Kaschke, M
机构
[1] UNIV ILLINOIS,DEPT PHYS,FLUORESCENCE DYNAM LAB,URBANA,IL 61801
[2] CARL ZEISS,MED OPT INSTRUMENTS,D-73446 OBERKOCHEN,GERMANY
[3] HUMBOLDT UNIV BERLIN,MAX DELBRUCK CTR MOL MED,ROBERT ROESSLE HOSP & TUMOR INST,D-13122 BERLIN,GERMANY
关键词
D O I
10.1073/pnas.94.12.6468
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
We present a novel approach to optical mammography and initial clinical results. We have designed and developed a frequency-domain (110-MHz) optical scanner that performs a transillumination raster scan of the female breast in approximately 3 min, The probing light is a dual-wavelength (690 and 810 nm, 10-mW average power), 2-mm-diameter laser beam, and the detection optical fiber is 5 mm in diameter. The ac amplitude and phase data are processed with use of an algorithm that performs edge effect corrections, thereby enhancing image contrast. This contrast enhancement results in a greater tumor detectability compared with simple light intensity images, The optical mammograms are displayed on a computer screen in real time, We present x-ray and optical mammograms from two patients with breast tumors. Our initial clinical results show that the frequency-domain scanner, even at the present stage of development, has the potential to be a useful tool in mammography.
引用
收藏
页码:6468 / 6473
页数:6
相关论文
共 23 条
[1]  
ALVERYD A, 1990, CANCER, V65, P1671, DOI 10.1002/1097-0142(19900415)65:8<1671::AID-CNCR2820650802>3.0.CO
[2]  
2-A
[3]   TIME-RESOLVED TRANSILLUMINATION FOR MEDICAL DIAGNOSTICS [J].
ANDERSSONENGELS, S ;
BERG, R ;
SVANBERG, S ;
JARLMAN, O .
OPTICS LETTERS, 1990, 15 (21) :1179-1181
[4]   TRANS-ILLUMINATION LIGHTSCANNING TO DIAGNOSE BREAST-CANCER - A FEASIBILITY STUDY [J].
BARTRUM, RJ ;
CROW, HC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 142 (02) :409-414
[5]   OPTICAL TIME-OF-FLIGHT AND ABSORBENCY IMAGING OF BIOLOGIC MEDIA [J].
BENARON, DA ;
STEVENSON, DK .
SCIENCE, 1993, 259 (5100) :1463-1466
[6]  
BENARON DA, 1995, P SOC PHOTO-OPT INS, V2389, P582
[7]  
Carlsen EN, 1982, DIAGN IMAGING, V4, P28
[8]  
Cutler M, 1929, Surg. Gynaecol, V48, P721
[9]   DIAPHANOGRAPHY IN THE DIAGNOSIS OF BREAST-CANCER [J].
DREXLER, B ;
DAVIS, JL ;
SCHOFIELD, G .
RADIOLOGY, 1985, 157 (01) :41-44
[10]   SPECTRAL TRANSMITTANCE AND CONTRAST IN BREAST DIAPHANOGRAPHY [J].
ERTEFAI, S ;
PROFIO, AE .
MEDICAL PHYSICS, 1985, 12 (04) :393-400