Feasibility of shock advice analysis during CPR through removal of CPR artefacts from the human ECG

被引:86
作者
Eilevstjonn, J
Eftestol, T
Aase, SO
Myklebust, H
Husoy, JH
Steen, PA
机构
[1] Stavanger Univ Coll, Dept Elect & Comp Engn, N-4068 Stavanger, Norway
[2] Laerdal Med, N-4002 Stavanger, Norway
[3] Ullevaal Univ Hosp, Dept Anesthesiol, N-0407 Oslo, Norway
关键词
out-of-hospital CPR; automated external defibrillators (AED); heart arrest; cardiac arrest; arrhythmia; electrocardiography; CPR artefact filtering;
D O I
10.1016/j.resuscitation.2003.12.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Mechanical activity from chest compressions and ventilations during cardiopulmonary resuscitation (CPR) introduces artefact components into the electrocardiogram (ECG). CPR must therefore be discontinued for reliable shock advice analysis in automated external defibrillators. Reducing or eliminating this detrimental "hands-off"timeby removing the CPR artefacts, should significantly improve the defibrillation success rate. The feasibility of this was tested by removing the CPR artefacts using a multichannel adaptive filter, the multichannel recursive adaptive matching pursuit (MC-RAMP) algorithm. Human ECG and reference channel data from episodes with both shockable and non-shockable underlying heart rhythms were recorded from 105 patients with out-of-hospital cardiac arrest. The performance of a shock advice algorithm was evaluated before and after artefact removal using the MC-RAMP algorithm. From a test set consisting of 92 shockable and 174 non-shockable episodes a sensitivity of 96.7% and specificity of 79.9% was achieved, an increase of approximately 15 and 13%, respectively, compared to no filtering. Good sensitivity was achieved, enabling ECG analysis during CPR that would reduce the hands-off time on patients with shockable rhythms. However, CPR artefact removal on non-shockable rhythms proved a more difficult problem. We need a better understanding of the physiological mixing of artefacts and the underlying heart rhythm and suggest clinical trials to investigate the nature of CPR artefacts further. (C) 2004 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:131 / 141
页数:11
相关论文
共 17 条
[1]   CPR artifact removal from human ECG using optimal multichannel filtering [J].
Aase, SO ;
Eftestol, T ;
Husoy, JH ;
Sunde, K ;
Steen, PA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2000, 47 (11) :1440-1449
[2]   Compression depth estimation for CPR quality assessment using DSP on accelerometer signals [J].
Aase, SO ;
Myklebust, H .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (03) :263-268
[3]  
*AG TECHN, 2000, AED ALG M3500B HEART
[4]   Effects of interrupting precordial compressions on the calculated probability of defibrillation success during out-of-hospital cardiac arrest [J].
Eftestol, T ;
Sunde, K ;
Steen, PA .
CIRCULATION, 2002, 105 (19) :2270-2273
[5]   Determination of the noise source in the electrocardiogram during cardiopulmonary resuscitation [J].
Fitzgibbon, E ;
Berger, R ;
Tsitlik, J ;
Halperin, HR .
CRITICAL CARE MEDICINE, 2002, 30 (04) :S148-S153
[6]  
Hart, 2006, PATTERN CLASSIFICATI
[7]   Removal of cardiopulmonary resuscitation artifacts from human ECG using an efficient matching pursuit-like algorithm [J].
Husoy, JH ;
Eilevstjonn, J ;
Eftestol, T ;
Aase, SO ;
Myklebust, H ;
Steen, PA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (11) :1287-1298
[8]   Reducing CPR artefacts in ventricular fibrillation in vitro [J].
Langhelle, A ;
Eftestol, T ;
Myklebust, H ;
Eriksen, M ;
Holten, BT ;
Steen, PA .
RESUSCITATION, 2001, 48 (03) :279-291
[9]   PREDICTING SURVIVAL FROM OUT-OF-HOSPITAL CARDIAC-ARREST - A GRAPHIC MODEL [J].
LARSEN, MP ;
EISENBERG, MS ;
CUMMINS, RO ;
HALLSTROM, AP .
ANNALS OF EMERGENCY MEDICINE, 1993, 22 (11) :1652-1658
[10]   ASSESSMENT OF THE VENTRICULAR-FIBRILLATION DETECTION ALGORITHM IN THE SEMIAUTOMATIC CARDIO-AID DEFIBRILLATOR [J].
MURRAY, A ;
CLAYTON, RH ;
CAMPBELL, RWF .
RESUSCITATION, 1995, 29 (02) :113-117