Traceable calibration of ultraviolet meters used with broadband, extended sources

被引:15
作者
Coleman, AJ [1 ]
Collins, M [1 ]
Saunders, JE [1 ]
机构
[1] St Thomas Hosp, Med Phys Directorate, London SE1 7EH, England
关键词
D O I
10.1088/0031-9155/45/1/313
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A calibration system has been developed to provide increased accuracy in the measurement of the irradiance responsivity appropriate for UV meters used with broadband, extended sources of the type employed in phototherapy. The single wavelength responsivity of the test meter is obtained in the wavelength range 250-400 nm by intercomparison with a transfer standard meter in a narrow, monochromatic beam. Traceability to primary standard irradiance scales is provided via the National Measurement System with a best uncertainty of 7% (at 95% confidence). The effective responsivity of the test meter, when used with broadband extended sources, is calculated using the measured spectral and angular response of the meter and tabulated data on the spectral and spatial characteristics of the source radiance. The uncertainty in the effective responsivity, independent of the source variability, is estimated to be 10% (at 95% confidence). The advantages of this calibration system over existing approaches are discussed.
引用
收藏
页码:185 / 196
页数:12
相关论文
共 21 条
[1]   STATUS OF ULTRAVIOLET-A DOSIMETRY IN METHOXSALEN PLUS ULTRAVIOLET-A THERAPY [J].
DIFFEY, BL ;
ROELANDTS, R .
JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 1986, 15 (06) :1209-1213
[2]   SURVEY OF THE ULTRAVIOLET-RADIATION EMISSIONS OF PHOTOCHEMOTHERAPY UNITS [J].
DIFFEY, BL ;
CHALLONER, AVJ ;
KEY, PJ .
BRITISH JOURNAL OF DERMATOLOGY, 1980, 102 (03) :301-306
[3]   PUVA - REVIEW OF ULTRAVIOLET DOSIMETRY [J].
DIFFEY, BL .
BRITISH JOURNAL OF DERMATOLOGY, 1978, 98 (06) :703-705
[4]  
DIFFEY BL, 1997, 76 IPEM, P25
[5]   Evaluation of solid-state detectors for ultraviolet radiometric applications [J].
Durant, NM ;
Fox, NP .
METROLOGIA, 1996, 32 (06) :505-508
[6]   High-accuracy characterisation and applications of filter radiometers [J].
Fox, NP ;
Chunnilall, CJ ;
Harrison, NJ ;
Hartree, WS .
OPTICAL RADIATION MEASUREMENTS III, 1996, 2815 :32-41
[7]  
FU L, 1993, METROLOGIA, V30, P297
[8]  
GOODMAN TM, 1989, BIOL TECHNIQUE SERIE, P47
[9]  
HANSEN AB, 1994, PHOTODERMATOL PHOTO, V10, P22
[10]  
HOLIDAKI E, 1995, THESIS U SURREY