Evaluation of pressure changes in a new design tracheal tube cuff, the Porter Soft Seal, during nitrous oxide anaesthesia

被引:26
作者
Al-Shaikh, B [1 ]
Jones, M [1 ]
Baldwin, F [1 ]
机构
[1] William Harvey Hosp, Dept Anaesthesia, Ashford TN24 0LZ, Kent, England
关键词
anaesthetics gases; nitrous oxide; equipment; cuffs tracheal; intubation tracheal;
D O I
10.1093/bja/83.5.805
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have measured pressure changes in a newly designed tracheal tube cuff, the Porter Soft Seal, during nitrous oxide anaesthesia compared with a Mallinckrodt Lo-Contour tube and a Porter Profile tube. The pressure increases in both control groups were significantly greater than those with the new design (P<0.0001 in each case). The mean increase in pressure in the Mallinckrodt Lo-Contour tube cuff was 9.9 (SD 3.4) mm Hg compared with 10.3 (1.8) mm Hg in the Porter Profile tube cuff and 2.1 (1.5) mm Hg in the Porter Soft Seal tube cuff. We conclude that the Porter Soft Seal cuff prevented a significant increase in intracuff pressure during nitrous oxide anaesthesia.
引用
收藏
页码:805 / 806
页数:2
相关论文
共 7 条
[1]  
[Anonymous], ACTA OTOLARYNGO S345
[2]  
Brandt L, 1990, REDUCTION NITROUS OX
[3]   PRESSURE CHANGES IN TRACHEAL TUBE CUFFS [J].
CHANDLER, M .
ANAESTHESIA, 1986, 41 (03) :287-293
[4]   Choice of cuff inflation medium during nitrous oxide anaesthesia [J].
Mitchell, V ;
Adams, T ;
Calder, I .
ANAESTHESIA, 1999, 54 (01) :32-36
[5]   THE MILLER TRACHEAL CUFF PRESSURE CONTROL VALVE - CLINICAL USE IN CONTROLLED AND SPONTANEOUS VENTILATION [J].
PAYNE, KA ;
MILLER, DM .
ANAESTHESIA, 1993, 48 (04) :324-327
[6]  
QUARTARARO C, 1990, SEMIN ANESTH, V9, P119
[7]  
RAEDER JC, 1985, ANAESTHESIA, V40, P444