Methods for improving the sensitivity in atom trapping flame atomic absorption spectrometry: Analytical scheme for the direct determination of trace elements in beer

被引:38
作者
Matusiewicz, H
Kopras, M
机构
[1] Politechnika Poznańska, Department of Analytical Chemistry
关键词
trace element determination; integrated atom trap system; flame atomic absorption spectrometry; sensitivity increase; beer;
D O I
10.1039/a704407k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method is described for the atomic absorption (AA) determination of Ag, Cd, Cu, Fe, In, Mn, Pb, Tl and Zn in beer using an 'integrated atom trap' system mounted on a standard AA air-acetylene flame burner, A new design of atom trapping technique that would exceed the operational capabilities of existing arrangements (a mater-cooled single or dual silica tube or a double-slotted quartz tube) and permit construction of an 'integrated trap' was investigated, A significant improvement in detection limit was achieved compared with that obtained using either of the above atom trapping techniques separately, Rapid, accurate analyses can be achieved using continuous aspiration, The concentration detection limits were 3.0, 5.0, 1.2, 4.0 and 0.1 ng ml(-1) for Cu, Fe, Mn, Pb and Zn, respectively, using a 2 min in situ preconcentration time, The relative standard deviations are of the order of 3.0-6.0% for this technique, Basic analytical performance characteristics are also given for Ag, Cd, In and Tl using various designs of atom trap, The designs studied include both slotted tube and single silica tube water-cooled atom traps.
引用
收藏
页码:1287 / 1291
页数:5
相关论文
共 22 条
[1]  
BORRIELLO R, 1980, ATOM SPECTROSC, V1, P131
[2]  
CERVERA L, 1989, J ASSOC OFF ANA CHEM, V72, P282
[3]   PLATFORM IN FURNACE ZEEMAN-EFFECT ATOMIC-ABSORPTION SPECTROMETRIC DETERMINATION OF ARSENIC IN BEER BY ATOMIZATION OF SLURRIES OF SAMPLE ASH [J].
CERVERA, ML ;
NAVARRO, A ;
MONTORO, R ;
DELAGUARDIA, M ;
SALVADOR, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (06) :477-481
[4]   Novel method to determine mercury in sediment using a gold coated dual 'bent' tube atom trap [J].
Ellis, LA ;
Roberts, DJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (11) :1063-1066
[5]   Further development of the 'bent' tube water cooled atom trap for arsenic, antimony, copper and manganese [J].
Ellis, LA ;
Roberts, DJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (04) :259-263
[6]  
HERGENREDER RI, 1991, ATOM SPECTROSC, V12, P74
[7]  
LAU C, 1976, CAN J SPECTROSC, V21, P100
[8]  
LEE M, 1985, BREWERS GUARDIAN, P13
[9]  
LI H, 1991, FENXI CESHI TONGBAO, V10, P51
[10]   PRECONCENTRATION OF TRACE-ELEMENTS IN POTABLE LIQUIDS BY MEAS OF A LIQUID MEMBRANE EMULSION FOR FLAME ATOMIC-ABSORPTION DETERMINATION [J].
LI, YT ;
VANLOON, JC ;
BAREFOOT, RR .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 345 (06) :467-470