Biomechanical evaluation of Caspar and Cervical Spine Locking Plate systems in a cadaveric model

被引:55
作者
Clausen, JD
Ryken, TC
Traynelis, VC
Sawin, PD
Dexter, F
Goel, VK
机构
[1] UNIV IOWA,DEPT BIOMED ENGN,DIV NEUROSURG,IOWA CITY,IA 52242
[2] UNIV IOWA,DEPT ANESTHESIA,IOWA CITY,IA 52242
关键词
biomechanics; cervical spine; fatigue loading; spinal fusion; spinal injury; spinal instrumentation;
D O I
10.3171/jns.1996.84.6.1039
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
There exist two markedly different instrumentation systems for the anterior cervical spine: the Cervical Spine Locking Plate (CSLP) system, which uses unicortical screws with a locking hub mechanism for attachment, and the Caspar Trapezial Plate System, which is secured with unlocked bicortical screws. The biomechanical stability of these two systems was evaluated in a cadaveric model of complete C5-6 instability. The immediate stability was determined in six loading modalities: flexion, extension, right and left lateral bending, and right and left axial rotation. Biomechanical stability was reassessed following fatigue with 5000 cycles of flexion-extension, and finally, the spines were loaded in flexion until the instrumentation failed. The Caspar system stabilized significantly in flexion before (p < 0.05) but not after fatigue, and it stabilized significantly in extension before (p < 0.01) and after fatigue (p < 0.01). The CSLP system stabilized significantly in flexion before (p < 0.01) but not after fatigue, and it did not stabilize in extension before or after fatigue. The moment needed to produce failure in flexion did not differ substantially between the two plating systems. The discrepancy in the biomechanical stability of these two systems may be due to differences in bone screw fixation.
引用
收藏
页码:1039 / 1045
页数:7
相关论文
共 21 条
  • [1] ANTERIOR CERVICAL FUSION AND CASPAR PLATE STABILIZATION FOR CERVICAL TRAUMA
    CASPAR, W
    BARBIER, DD
    KLARA, PM
    [J]. NEUROSURGERY, 1989, 25 (04) : 491 - 502
  • [2] CASPAR W, 1991, MICROSURGERY CERVICA, P109
  • [3] GALLAGHER MR, 1993, NEUROSURGERY, V33, P1045
  • [4] KINEMATICS OF THE WHOLE LUMBAR SPINE - EFFECT OF DISCECTOMY
    GOEL, VK
    GOYAL, S
    CLARK, C
    NISHIYAMA, K
    NYE, T
    [J]. SPINE, 1985, 10 (06) : 543 - 554
  • [5] RESPONSE OF THE LIGAMENTOUS LUMBAR SPINE TO CYCLIC BENDING LOADS
    GOEL, VK
    VOO, LM
    WEINSTEIN, JN
    LIU, YK
    OKUMA, T
    NJUS, GO
    [J]. SPINE, 1988, 13 (03) : 294 - 300
  • [6] A TECHNIQUE TO EVALUATE AN INTERNAL SPINAL DEVICE BY USE OF THE SELSPOT SYSTEM - AN APPLICATION TO LUQUE CLOSED-LOOP
    GOEL, VK
    NYE, TA
    CLARK, CR
    NISHIYAMA, K
    WEINSTEIN, JN
    [J]. SPINE, 1987, 12 (02) : 150 - 159
  • [7] LABORATORY CONDITIONS FOR TENSILE TESTS IN ANNULUS FIBROSUS FROM HUMAN INTERVERTEBRAL DISCS
    HIRSCH, C
    GALANTE, J
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1967, 38 (02): : 148 - &
  • [8] FATIGUE RESPONSE OF LUMBAR INTERVERTEBRAL JOINTS UNDER AXIAL CYCLIC LOADING
    LIU, YK
    NJUS, G
    BUCKWALTER, J
    WAKANO, K
    [J]. SPINE, 1983, 8 (08) : 857 - 865
  • [9] MANN RW, 1983, B HOSP JOINT DIS ORT, V43, P137
  • [10] BIOMECHANICAL TIME-TOLERANCE OF FRESH CADAVERIC HUMAN SPINE SPECIMENS
    PANJABI, MM
    KRAG, M
    SUMMERS, D
    VIDEMAN, T
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1985, 3 (03) : 292 - 300