Estimating plasma volume from red cell volume

被引:4
作者
Cosgriff, PS [1 ]
Blunkett, M [1 ]
Morrish, O [1 ]
机构
[1] Pilgrim Hosp, Dept Med Phys, Boston PE21 9QS, Lincs, England
关键词
D O I
10.1097/00006231-199912000-00010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Following the withdrawal of the only licensed supply of I-125-HSA in 1997, most UK centres now simply estimate plasma volume from a knowledge of the red cell volume and venous haematocrit. We compared measured and estimated plasma volume in 107 consecutive patients who had had red cell and plasma volume measured independently in the conventional way. In only 40% of cases was the estimated value accurate to within +/- 5%, and discrepancies of more than +/- 10% were seen in 35%. As there appears to be no simple way of predicting those patients for whom an estimation is inappropriate, it is recommended that plasma volume is measured independently in all patients, using a reputable source of unlicensed I-125-HSA. ((C) 1999 Lippincott Williams & Wilkins).