Diode laser induced near infra-red fluorescence detection for high-performance liquid chromatography analysis of indocyanine green in plasma

被引:7
作者
Gui, M [1 ]
Nagaraj, S [1 ]
Rahavendran, SV [1 ]
Karnes, HT [1 ]
机构
[1] VIRGINIA COMMONWEALTH UNIV,MED COLL VIRGINIA,DEPT PHARM & PHARMACEUT,RICHMOND,VA 23298
关键词
lasers; fluorimetry; high-performance liquid chromatography; infrared spectrometry; biopharmaceutical analysis; indocyanine green; plasma;
D O I
10.1016/S0003-2670(96)00574-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Diode laser induced near infra-red fluorescence (DLINIRF) detection has become important in bioanalysis due to the increased sensitivity and selectivity that can be achieved in biological matrices as well as the ruggedness, low cost, stability and portability of diode lasers. Indocyanine green (ICG) is used as a marker in clinical research to measure hepatic blood how (HBF) and cardiac output. A high-performance liquid chromatography (HPLC) method employing a simple protein precipitation sample preparation step and DLINIRF detection has been developed to quantitate ICG in plasma. Comparison with absorbance detection demonstrated an increased detectability of more than two orders of magnitude for the DLINIRF detection method. The analysis of ICG in plasma by both detection methods was selective and demonstrated accuracy and precision within acceptable limits.
引用
收藏
页码:145 / 149
页数:5
相关论文
共 12 条
[1]  
AWNL WM, 1989, CLIN CHEM, V35, P2124
[2]   LACK OF EFFECT OF HISTAMINE H-2-RECEPTOR ANTAGONISTS ON INDOCYANINE GREEN DISPOSITION MEASURED BY 2 METHODS [J].
DONN, KH ;
POWELL, JR ;
ROGERS, JF ;
PLACHETKA, JR .
JOURNAL OF CLINICAL PHARMACOLOGY, 1984, 24 (8-9) :360-370
[3]   SPECIFIC ASSAY FOR THE QUANTITATION OF INDOCYANINE GREEN IN RAT PLASMA USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE DETECTION [J].
DORR, MB ;
POLLACK, GM .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1989, 78 (04) :328-333
[4]   EXPERIMENTAL IMPACT OF ASSAY-DEPENDENT DIFFERENCES IN PLASMA INDOCYANINE GREEN CONCENTRATION DETERMINATIONS [J].
GRASELA, DM ;
ROCCI, ML ;
VLASSES, PH .
JOURNAL OF PHARMACOKINETICS AND BIOPHARMACEUTICS, 1987, 15 (06) :601-613
[5]   EVALUATION OF A NEW CONTINUOUS THERMODILUTION CARDIAC-OUTPUT MONITOR IN CRITICALLY ILL PATIENTS - A PROSPECTIVE CRITERION STANDARD STUDY [J].
HALLER, M ;
ZOLLNER, C ;
BRIEGEL, J ;
FORST, H .
CRITICAL CARE MEDICINE, 1995, 23 (05) :860-866
[6]  
HEILMEYER L, 1943, SPECTROPHOTOMETRY ME
[7]  
HOLLINS B, 1987, CLIN CHEM, V33, P765
[8]  
KESSEL D, 1984, METHODS PORPHYRIN PH
[9]   STABILITY OF INDOCYANINE GREEN IN HUMAN SERUM STORED AT -20-DEGREES-C AND -70-DEGREES-C [J].
MARCH, C ;
ADAMS, MH ;
GARNETT, W ;
KARNES, HT .
THERAPEUTIC DRUG MONITORING, 1994, 16 (06) :588-591
[10]   COMPARISON OF THE SPECTROPHOTOMETRIC AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS OF INDOCYANINE GREEN IN ESTIMATING SYSTEMIC CLEARANCE IN PATIENTS WITH CHRONIC LIVER-DISEASE [J].
OREILLY, T ;
MACMATHUNA, P ;
KEELING, PWN ;
FEELY, J .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 417 (01) :190-196