System peaks in capillary zone electrophoresis .3. Practical rules for predicting the existence of system peaks in capillary zone electrophoresis of anions using indirect spectrophotometric detection

被引:44
作者
Macka, M
Haddad, PR
Gebauer, P
Bocek, P
机构
[1] ACAD SCI CZECH REPUBL,INST ANALYT CHEM,CZ-61142 BRNO,CZECH REPUBLIC
[2] UNIV TASMANIA,DEPT CHEM,SEPARAT SCI GRP,HOBART,TAS 7001,AUSTRALIA
关键词
system peaks; indirect detection; capillary zone electrophoresis;
D O I
10.1002/elps.1150181120
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A theoretical and experimental study of the existence and evolution of system peaks in capillary zone electrophoresis (CZE) with indirect spectrophotometric detection is presented with respect to the effect of the number of colons present in the background electrolyte (BGE). It is shown that in BGEs having only one colon (i.e., the UV-absorbing probe anion), the sample produces only negative peaks due to each analyte anion and no system peaks, with the number of sample peaks corresponding to the number of analytes present in the sample injected. In BGEs containing two colons, a sample with one analyte anion produces one negative indirect detection peak and one system peak. The transition between BGEs having one colon and those with two colons has also been studied and it has been shown that an addition of ca. 5% of the second colon to a single colon BGE causes the resulting BGE to behave macroscopically as a regular two-colon BGE. A descriptive model is proposed, based on transient isotachophoresis (transient stacking) of the sample species and of the colon from the BGE which has the closest mobility to the sample ion. This model explains qualitatively the formation and evolution of the sample peak (containing the sample species and being detected by indirect detection due to displacement of the UV-absorbing probe in its zone) and the system peak (containing no sample species and being a vacancy in the continuum of colons of the BGE). It is shown that the system peak may be positive or negative as it corresponds to the situation where the vacancy of one component of the BGE results in an enhanced concentration of the other component. It has been demonstrated that the system peak is created by a vacancy of that component of the BGE which has the greatest difference in mobility relative to that of the sample species. On indirect detection in BGEs containing two colons the sample displaces predominantly the BGE colon which has a mobility closest to that of the analyte anion. In systems with BGEs containing two colons, a sample having n analytes produces n sample peaks and one system peak, the sign and magnitude of which are dependent on the sum of the UV absorbances of the analytes involved. The effect of bicarbonate from atmospheric CO2 has also been studied and it has been shown that weakly alkaline BGEs with a single anionic UV-absorbing colon, such as those currently used for anionic analyses with indirect detection, may suffer from the presence of system peaks due to bicarbonate.
引用
收藏
页码:1998 / 2007
页数:10
相关论文
共 22 条
[1]   SYSTEM PEAKS AND DISTURBANCES TO THE BASE-LINE UV SIGNAL IN CAPILLARY ZONE ELECTROPHORESIS [J].
BECKERS, JL .
JOURNAL OF CHROMATOGRAPHY A, 1994, 662 (01) :153-166
[2]   CALCULATION OF THE COMPOSITION OF SAMPLE ZONES IN CAPILLARY ZONE ELECTROPHORESIS .2. SIMULATED ELECTROPHEROGRAMS [J].
BECKERS, JL .
JOURNAL OF CHROMATOGRAPHY A, 1995, 696 (02) :285-294
[3]   UV DETECTION IN CAPILLARY ZONE ELECTROPHORESIS PEAKS OR DIPS - THAT IS THE QUESTION [J].
BECKERS, JL .
JOURNAL OF CHROMATOGRAPHY A, 1994, 679 (01) :153-165
[4]  
Bocek P., 1988, ANAL ISOTACHOPHORESI
[5]   THEORETICAL AND EXPERIMENTAL ASPECTS OF INDIRECT DETECTION IN CAPILLARY ELECTROPHORESIS [J].
BRUIN, GJM ;
VANASTEN, AC ;
XU, XM ;
POPPE, H .
JOURNAL OF CHROMATOGRAPHY, 1992, 608 (1-2) :97-107
[6]   System peaks in capillary zone electrophoresis .2. Experimental study of vacancy peaks [J].
Desiderio, C ;
Fanali, S ;
Gebauer, P ;
Bocek, P .
JOURNAL OF CHROMATOGRAPHY A, 1997, 772 (1-2) :81-89
[7]  
DOBLE P, 1996, INT ION CHROM S 96 R
[8]  
Foret F., 1993, CAPILLARY ZONE ELECT
[9]   System peaks in capillary zone electrophoresis .1. Simple model of vacancy electrophoresis [J].
Gebauer, P ;
Bocek, P .
JOURNAL OF CHROMATOGRAPHY A, 1997, 772 (1-2) :73-79
[10]  
GEBAUER P, 1997, 9 INT S HIGH PERF CA, P70