Speciation of arsenic compounds in drinking water by capillary electrophoresis with hydrodynamically modified electroosmotic flow detected through hydride generation inductively coupled plasma mass spectrometry with a membrane gas-liquid separator

被引:89
作者
Magnuson, ML [1 ]
Creed, JT [1 ]
Brockhoff, CA [1 ]
机构
[1] US EPA,NATL EXPOSURE RES LAB,HUMAN EXPOSURE RES DIV,CHEM EXPOSURE RES BRANCH,CINCINNATI,OH 45268
关键词
speciation; arsenic compounds; capillary electrophoresis; electrokinetic injection; modified electroosmotic flow; hydride generation; microporous membrane gas-liquid separator; inductively coupled plasma mass spectrometry;
D O I
10.1039/a607730g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
environmentally significant, toxic forms of arsenic: arsenite, arsenate, monomethylarsonic acid and dimethylarsinic acid, Hydride generation (HG) was used to convert the species into their respective hydrides, The hydride species were detected with inductively coupled plasma mass spectrometry, The HG unit utilized a microporous PTFE tube as a gas-liquid separator, The injection mode for CE was electrokinetic in conjunction with the novel use of hydrodynamically modified electroosmotic flow (HMEOF). In HMEOF, the electroosmotic flow is modified by applying hydrodynamic pressure opposite to the direction of the electroosmotic flow. HMEOF provides the capability of injecting increased quantities of analyte by offsetting the electroosmotic flow, which limits conventional electrokinetic injection, In order to correct for imprecisions in the electrokinetic injection in matrices of different ionic strength, the use of a surrogate for the injection of arsenic species was investigated, Germanium was investigated because it forms a hydride and has a low natural occurrence, The separation also utilized HMEOF, which allowed for greater freedom in buffer choice, The detection limits in distilled, de-ionized mater were 25, 6, 9 and 58 ppt for the four species listed above, respectively, The detection limit was calculated from 3.14 sigma(n-1) of seven replicate injections and represents the precision of measuring the ratio of the area of the arsenic peaks to the area of a germanium surrogate peak, Standard addition was used to determine arsenate in drinking water samples, Recoveries of arsenite and arsenate from drinking water samples are reported using germanium as a surrogate to correct for sampling bias of the electrokinetic injection.
引用
收藏
页码:689 / 695
页数:7
相关论文
共 49 条
[1]   ARSENIC SPECIATION IN MARINE BIOLOGICAL-MATERIALS BY LC-UV-HG-ICP/OES [J].
ALBERTI, J ;
RUBIO, R ;
RAURET, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1995, 351 (4-5) :415-419
[2]   MICROPOROUS POLYTETRAFLUOROETHYLENE TUBING GAS-LIQUID SEPARATOR FOR HYDRIDE GENERATION AND INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BARNES, RM ;
WANG, XR .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (08) :1083-1089
[3]   DETERMINATION OF ARSENIC SPECIES BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY - INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
SIU, KWM ;
MCLAREN, JW ;
BERMAN, SS .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (03) :285-289
[4]   DETERMINATION OF ARSENIC BY HYDRIDE GENERATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY USING A TUBULAR MEMBRANE GAS-LIQUID SEPARATOR [J].
BRANCH, S ;
CORNS, WT ;
EBDON, L ;
HILL, S ;
ONEILL, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (02) :155-158
[5]   DETERMINATION OF ARSENIC SPECIES IN FISH BY DIRECTLY COUPLED HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY-INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY [J].
BRANCH, S ;
EBDON, L ;
ONEILL, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (01) :33-37
[6]   NEW CONCEPT FOR HYDRIDE GENERATION TECHNIQUE - ELECTROCHEMICAL HYDRIDE GENERATION [J].
BROCKMANN, A ;
NONN, C ;
GOLLOCH, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (03) :397-401
[7]   DETERMINATION OF SELENIUM BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY UTILIZING A NEW HYDRIDE GENERATION SAMPLE INTRODUCTION SYSTEM [J].
BUCKLEY, WT ;
BUDAC, JJ ;
GODFREY, DV ;
KOENIG, KM .
ANALYTICAL CHEMISTRY, 1992, 64 (07) :724-729
[9]   FIELD-AMPLIFIED POLARITY-SWITCHING SAMPLE INJECTION IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS [J].
CHIEN, RL ;
BURGI, DS .
JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2) :153-161
[10]   ON-COLUMN SAMPLE CONCENTRATION USING FIELD AMPLIFICATION IN CZE [J].
CHIEN, RL ;
BURGI, DS .
ANALYTICAL CHEMISTRY, 1992, 64 (08) :A489-A496