Clinical experience with a novel endotoxin adsorbtion device in patients undergoing cardiac surgery

被引:17
作者
Blomquist, S. [1 ]
Gustafsson, V. [2 ]
Manolopoulos, T. [1 ]
Pierre, L. [2 ]
机构
[1] Univ Lund Hosp, Dept Cardiothorac Anaesthesia, Ctr Heart & Lung Dis, S-22185 Lund, Sweden
[2] Univ Lund Hosp, Dept Perfus, Ctr Heart & Lung Dis, S-22185 Lund, Sweden
来源
PERFUSION-UK | 2009年 / 24卷 / 01期
关键词
adsorbtion; cardiopulmonary bypass; endotoxin; inflammatory mediators; IMMOBILIZED HEMOPERFUSION CARTRIDGE; COMPLEMENT; REMOVAL; SEPSIS;
D O I
10.1177/0267659109106730
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Endotoxaemia is thought to occur in cardiac surgery using extracorporeal circulation (ECC) and a positive correlation has been proposed between the magnitude of endotoxaemia and risk for postoperative complications. We studied the effects of a new endotoxin adsorber device (Alteco (R) LPS adsorber) in patients undergoing cardiac surgery with ECC, with special reference to safety and ease of use. Fifteen patients undergoing coronary artery bypass and/or valvular surgery were studied. In 9 patients, the LPS Adsorber was included in the bypass circuit between the arterial filter and the venous reservoir. Flow through the adsorber was started when the aorta was clamped and stopped at the end of perfusion. Flow rate was kept at 150 ml/min. Six patients served as controls with no adsorber in the circuit. Samples were taken for analysis of endotoxin, TNF alpha, IL- 1 beta and IL-6 as well as complement factors C3, C4 and C1q. Whole blood coagulation status was evaluated using thromboelastograpy (TEG) and platelet count. No adverse events were encountered when the adsorber was used in the circuit. Blood flow through the device was easily monitored and kept at the desired level. Platelet count decreased in both groups during surgery. TEG data revealed a decrease in whole blood clot strength in the control group while it was preserved in the adsorber group. Endotoxin was detected in only 2 patients and IL-1 beta in 4 patients. IL-6 decreased in both groups whereas no change in TNF concentrations was found. C3 fell in both groups, but no changes wer found in C4 and C1q. The Alteco (R) LPS adsorber can be used safely and is easy to handle in the bypass circuit. No complications related to the use of the adsorber were noted. The intended effects of the adsorber, i.e. removal of endotoxin from the blood stream could not be evaluated in this study, presumably due to the small number of patients and the relatively short perfusion times. Perfusion (2009) 24, 13-17.
引用
收藏
页码:13 / 17
页数:5
相关论文
共 14 条
[1]   COMPLEMENT IN ORGAN-TRANSPLANTATION - CONTRIBUTIONS TO INFLAMMATION, INJURY, AND REJECTION [J].
BALDWIN, WM ;
PRUITT, SK ;
BRAUER, RB ;
DAHA, MR ;
SANFILIPPO, F .
TRANSPLANTATION, 1995, 59 (06) :797-808
[2]   Therapeutic potential of targeting the complement cascade in critical care medicine [J].
Bhole, D ;
Stahl, GL .
CRITICAL CARE MEDICINE, 2003, 31 (01) :S97-S104
[3]   INFLAMMATORY RESPONSE TO CARDIOPULMONARY BYPASS [J].
BUTLER, J ;
ROCKER, GM ;
WESTABY, S .
ANNALS OF THORACIC SURGERY, 1993, 55 (02) :552-559
[4]   COMPLEMENT ACTIVATION DURING CARDIOPULMONARY BYPASS - EVIDENCE FOR GENERATION OF C3A AND C5A ANAPHYLATOXINS [J].
CHENOWETH, DE ;
COOPER, SW ;
HUGLI, TE ;
STEWART, RW ;
BLACKSTONE, EH ;
KIRKLIN, JW .
NEW ENGLAND JOURNAL OF MEDICINE, 1981, 304 (09) :497-503
[5]   ENDOTOXEMIA IN HUMAN SEPTIC SHOCK [J].
DANNER, RL ;
ELIN, RJ ;
HOSSEINI, JM ;
WESLEY, RA ;
REILLY, JM ;
PARILLO, JE .
CHEST, 1991, 99 (01) :169-175
[6]   Strategies for targeting complement inhibitors in ischaemia/reperfusion injury [J].
Dong, J ;
Pratt, JR ;
Smith, RAG ;
Dodd, I ;
Sacks, SH .
MOLECULAR IMMUNOLOGY, 1999, 36 (13-14) :957-963
[7]  
KIRKLIN JK, 1983, J THORAC CARDIOV SUR, V86, P845
[8]   Endotoxin adsorption therapy for septic shock using polymyxin B-immobilized fibers (PMX): Evaluation by high-sensitivity endotoxin assay and measurement of the cytokine production capacity [J].
Kojika, M ;
Sato, N ;
Yaegashi, Y ;
Suzuki, Y ;
Suzuki, K ;
Nakae, H ;
Endo, S .
THERAPEUTIC APHERESIS AND DIALYSIS, 2006, 10 (01) :12-18
[9]   Use of an endotoxin adsorber in the treatment of severe abdominal sepsis [J].
Kulabukhov, V. V. .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2008, 52 (07) :1024-1025
[10]  
LEVY JH, 2003, ANN THORAC SURG, V75, pS175