A Wearable System to Assist Walking of Parkinson's Disease Patients Benefits and Challenges of Context-triggered Acoustic Cueing

被引:109
作者
Baechlin, M. [1 ]
Plotnik, M. [2 ,3 ]
Roggen, D. [1 ]
Giladi, N. [2 ,4 ]
Hausdorff, J. M. [2 ,5 ,6 ]
Troester, G. [1 ]
机构
[1] ETH, Wearable Comp Lab, Swiss Fed Inst Technol, CH-8092 Zurich, Switzerland
[2] Tel Aviv Sourasky Med Ctr, Dept Neurol, Movement Disorders Unit, Lab Gait & Neurodynam, Tel Aviv, Israel
[3] Bar Ilan Univ, Gonda Brain Res Ctr, Ramat Gan, Israel
[4] Tel Aviv Univ, Sackler Fac Med, Dept Neurol, IL-69978 Tel Aviv, Israel
[5] Tel Aviv Univ, Sackler Fac Med, Dept Phys Therapy, IL-69978 Tel Aviv, Israel
[6] Harvard Univ, Sch Med, Div Aging, Boston, MA 02115 USA
关键词
Pervasive health; wearable system; Parkinson's disease; RHYTHMIC AUDITORY-STIMULATION; GAIT; TRIAL;
D O I
10.3414/ME09-02-0003
中图分类号
TP [自动化技术、计算机技术];
学科分类号
080201 [机械制造及其自动化];
摘要
Background: About 50% of the patients with advanced Parkinson's disease (PD) suffer from freezing of gait (FOG), which is a sudden and transient inability to walk. it often causes falls, interferes with daily activities and significantly impairs quality of life. Because gait deficits in PD patients are often resistant to pharmacologic treatment, effective non-pharmacologic treatments are of special interest. Objectives: The goal of our study is to evaluate the concept of a wearable device that can obtain real-time gait data, processes them and provides asistance based on pre-determined specifications. Methods: We developed a real-time wearable FOG detection system that automatically provides a cueing sound when FOG is detected and which stays until the subject resumes walking. We evaluated our wearable assistive technology in a study with 10 PD patients. Over eight hours of data was recorded and a questionnaire was filled out by each patient. Results: Two hundred and thirty-seven FOG events have been identified by professional physiotherapists in a post-hoc video analysis, The device detected the FOG events online with a sensitivity of 73.1% and a specificity of 81.6% on a 0.5 sec frame-based evaluation. Conclusions: With this study we show that online assistive feedback for PD patients is possible. We present and discuss the patients' and physiotherapists' perspectives on wearabilty and performance of the wearable assistant as well as their gait performance when using the assistant and point out the next research steps. our results demonstrate the benefit of such a context-aware system and motivate further studies.
引用
收藏
页码:88 / 95
页数:8
相关论文
共 24 条
[1]
AMFT O, 2004, ASAP 04, P398
[2]
[Anonymous], 2009, P 4 INT C BOD AR NET
[3]
BACHLIN M, 2007, P 2 SENSATION INT C
[4]
BACHLIN M, 2009, P 13 INT S WEAR COMP, P123
[5]
BACHLIN M, 2009, P 3 INT C PERV COMP
[6]
Falls and freezing of gait in Parkinson's disease: A review of two interconnected, episodic phenomena [J].
Bloem, BR ;
Hausdorff, JA ;
Visser, JE ;
Giladi, N .
MOVEMENT DISORDERS, 2004, 19 (08) :871-884
[7]
Stages in the development of Parkinson's disease-related pathology [J].
Braak, H ;
Ghebremedhin, E ;
Rüb, U ;
Bratzke, H ;
Del Tredici, K .
CELL AND TISSUE RESEARCH, 2004, 318 (01) :121-134
[8]
Short-term and practice effects of metronome pacing in Parkinson's disease patients with gait freezing while in the 'on' state: randomized single blind evaluation [J].
Cubo, E ;
Leurgans, S ;
Goetz, CG .
PARKINSONISM & RELATED DISORDERS, 2004, 10 (08) :507-510
[9]
Quality of life in patients with Parkinson's disease: Development of a questionnaire [J].
deBoer, AGEM ;
Wijker, W ;
Speelman, JD ;
deHaes, JCJM .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1996, 61 (01) :70-74
[10]
FAHN S, 1995, ADV NEUROL, V67, P53