Evaluation of the Critikon 8100 and Spacelabs 90207 non-invasive blood pressure monitors using a test simulator

被引:18
作者
Amoore, JN
Geake, WB
机构
[1] Dept. of Med. Phys. and Med. Eng., Royal Infirmary of Edinburgh, Edinburgh
[2] Dept. of Med. Phys. and Med. Eng., Royal Infirmary of Edinburgh, Edinburgh EH3 9YW, Lauriston Place
关键词
NIBP monitors; evaluation; simulators;
D O I
10.1038/sj.jhh.1000416
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The Critikon Dinamap 8100 and the Spacelabs 90207 ambulatory non-invasive blood pressure (NIBP) monitors were evaluated using a test simulator using an evaluation protocol which covered a wide range of Simulated pressures (with five determinations at each of six steps from 60/30 to 200/150 mm Hg), pulse rates (from 40 to 200 bpm), artefact levels (simulated motion and tremor artefact) and pulse strengths (down to 10% of the nominal strength). Determinations were made at 5 min intervals. The average and standard deviation of the five measurements at each condition were calculated. The Spacelabs recorded pressures with a greater consistency than the Dinamap which showed a higher standard deviation under all the conditions. The relatively high standard deviation of the recordings made by the Dinamap could explain the non-systematic errors found in some evaluations. Both instruments recorded pressures within 5 mm Hg of the target over the range of pressures and pulse rates, and coped well under conditions of severe artefact and weak pulsations, either by signalling inability to record or by recording satisfactorily.
引用
收藏
页码:163 / 169
页数:7
相关论文
共 18 条
[1]  
Alpert, 1996, Blood Press Monit, V1, P69
[2]   Oscillometric non-invasive blood pressure measurements: The influence of the make of instrument on readings? [J].
Amoore, JN ;
Geake, WB ;
Scott, DHT .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (02) :131-134
[3]   ASSESSMENT OF OSCILLOMETRIC NONINVASIVE BLOOD-PRESSURE MONITORS USING THE DYNATECH NEVADA CUFFLINK ANALYZER [J].
AMOORE, JN .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1993, 17 (01) :25-31
[4]   An automated system for the functional evaluation of oscillometric non-invasive blood pressure monitors [J].
Geake, WB ;
Amoore, JN ;
Scott, DHT .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1995, 19 (05) :162-176
[5]   CARDIAC CYCLE PHASE UNCERTAINTY - ANOTHER SOURCE OF ERROR IN INDIRECT BLOOD-PRESSURE MEASUREMENT [J].
HATSELL, CP .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1992, 16 (04) :157-158
[6]  
MIEKE S, 1993, ANAESTHESIST, V42, P38
[7]  
Ng Kim-Gau, 1992, Journal of Clinical Engineering, V17, P469
[8]  
Ng Kim-Gau, 1994, Journal of Clinical Engineering, V19, P452
[9]  
OBRIEN E, 1994, J HUM HYPERTENS, V8, P73
[10]   ACCURACY OF THE SPACELABS-90207 DETERMINED BY THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL [J].
OBRIEN, E ;
MEE, F ;
ATKINS, N ;
OMALLEY, K .
JOURNAL OF HYPERTENSION, 1991, 9 (06) :573-574