Improved UTE-based attenuation correction for cranial PET-MR using dynamic magnetic field monitoring

被引:36
作者
Aitken, A. P. [1 ]
Giese, D. [1 ]
Tsoumpas, C. [1 ]
Schleyer, P. [1 ]
Kozerke, S. [1 ,2 ]
Prieto, C. [1 ]
Schaeffter, T. [1 ]
机构
[1] Kings Coll London, St Thomas Hosp, Kings Hlth Partners, Dept Biomed Engn,Div Imaging Sci & Biomed Engn, London SE1 7EH, England
[2] ETH, Inst Biomed Engn, CH-8092 Zurich, Switzerland
基金
英国惠康基金;
关键词
Magnetic resonance imaging; positron emission tomography; attenuation correction; ultrashort echo time; eddy currents; artifacts; magnetic field monitoring; ULTRASHORT ECHO-TIME; DISTORTION CORRECTION; MOTION COMPENSATION; PET/MRI; RECONSTRUCTION; SEGMENTATION; IMAGES; SYSTEMS; ATLAS; HEAD;
D O I
10.1118/1.4837315
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Ultrashort echo time (UTE) MRI has been proposed as a way to produce segmented attenuation maps for PET, as it provides contrast between bone, air, and soft tissue. However, UTE sequences require samples to be acquired during rapidly changing gradient fields, which makes the resulting images prone to eddy current artifacts. In this work it is demonstrated that this can lead to misclassification of tissues in segmented attenuation maps (AC maps) and that these effects can be corrected for by measuring the true k-space trajectories using a magnetic field camera. Methods: The k-space trajectories during a dual echo UTE sequence were measured using a dynamic magnetic field camera. UTE images were reconstructed using nominal trajectories and again using the measured trajectories. A numerical phantom was used to demonstrate the effect of reconstructing with incorrect trajectories. Images of an ovine leg phantom were reconstructed and segmented and the resulting attenuation maps were compared to a segmented map derived from a CT scan of the same phantom, using the Dice similarity measure. The feasibility of the proposed method was demonstrated in in vivo cranial imaging in five healthy volunteers. Simulated PET data were generated for one volunteer to show the impact of misclassifications on the PET reconstruction. Results: Images of the numerical phantom exhibited blurring and edge artifacts on the bone-tissue and air-tissue interfaces when nominal k-space trajectories were used, leading to misclassification of soft tissue as bone and misclassification of bone as air. Images of the tissue phantom and the in vivo cranial images exhibited the same artifacts. The artifacts were greatly reduced when the measured trajectories were used. For the tissue phantom, the Dice coefficient for bone in MR relative to CT was 0.616 using the nominal trajectories and 0.814 using the measured trajectories. The Dice coefficients for soft tissue were 0.933 and 0.934 for the nominal and measured cases, respectively. For air the corresponding figures were 0.991 and 0.993. Compared to an unattenuated reference image, the mean error in simulated PET uptake in the brain was 9.16% when AC maps derived from nominal trajectories was used, with errors in the SUVmax for simulated lesions in the range of 7.17%-12.19%. Corresponding figures when AC maps derived from measured trajectories were used were 0.34% (mean error) and -0.21% to + 1.81% (lesions). Conclusions: Eddy current artifacts in UTE imaging can be corrected for by measuring the true k-space trajectories during a calibration scan and using them in subsequent image reconstructions. This improves the accuracy of segmented PET attenuation maps derived from UTE sequences and subsequent PET reconstruction. (C) 2014 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution 3.0 Unported License.
引用
收藏
页数:13
相关论文
共 43 条
[1]   Characterization and Correction of System Delays and Eddy Currents for MR Imaging with Ultrashort Echo-Time and Time-Varying Gradients [J].
Atkinson, Ian C. ;
Lu, Aiming ;
Thulborn, Keith R. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (02) :532-537
[2]   A new improved version of the realistic digital brain phantom [J].
Aubert-Broche, Berengere ;
Evans, Alan C. ;
Collins, Louis .
NEUROIMAGE, 2006, 32 (01) :138-145
[3]   Spatiotemporal magnetic field monitoring for MR [J].
Barmet, Christoph ;
De Zanche, Nicola ;
Pruessmann, Klaas P. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (01) :187-197
[4]   A Transmit/Receive System for Magnetic Field Monitoring of In Vivo MRI [J].
Barmet, Christoph ;
De Zanche, Nicola ;
Wilm, Bertram J. ;
Pruessmann, Klaas P. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (01) :269-276
[5]   MRI-Based Attenuation Correction for Hybrid PET/MRI Systems: A 4-Class Tissue Segmentation Technique Using a Combined Ultrashort-Echo-Time/Dixon MRI Sequence [J].
Berker, Yannick ;
Franke, Jochen ;
Salomon, Andre ;
Palmowski, Moritz ;
Donker, Henk C. W. ;
Temur, Yavuz ;
Mottaghy, Felix M. ;
Kuhl, Christiane ;
Izquierdo-Garcia, David ;
Fayad, Zahi A. ;
Kiessling, Fabian ;
Schulz, Volkmar .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (05) :796-804
[6]   Investigation of MR-Based Attenuation Correction and Motion Compensation for Hybrid PET/MR [J].
Buerger, Christian ;
Tsoumpas, Charalampos ;
Aitken, Andrew ;
King, Andrew Peter ;
Schleyer, Paul ;
Schulz, Volkmar ;
Marsden, Paul K. ;
Schaeffter, Tobias .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (05) :1967-1976
[7]   Toward Implementing an MRI-Based PET Attenuation-Correction Method for Neurologic Studies on the MR-PET Brain Prototype [J].
Catana, Ciprian ;
van der Kouwe, Andre ;
Benner, Thomas ;
Michel, Christian J. ;
Hamm, Michael ;
Fenchel, Matthias ;
Fischl, Bruce ;
Rosen, Bruce ;
Schmand, Matthias ;
Sorensen, A. Gregory .
JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (09) :1431-1438
[8]   MEASURES OF THE AMOUNT OF ECOLOGIC ASSOCIATION BETWEEN SPECIES [J].
DICE, LR .
ECOLOGY, 1945, 26 (03) :297-302
[9]   A complete distortion correction for MR images: I. Gradient warp correction [J].
Doran, SJ ;
Charles-Edwards, L ;
Reinsberg, SA ;
Leach, MO .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (07) :1343-1361
[10]   First Clinical Experience with Integrated Whole-Body PET/MR: Comparison to PET/CT in Patients with Oncologic Diagnoses [J].
Drzezga, Alexander ;
Souvatzoglou, Michael ;
Eiber, Matthias ;
Beer, Ambros J. ;
Fuerst, Sebastian ;
Martinez-Moeller, Axel ;
Nekolla, Stephan G. ;
Ziegler, Sibylle ;
Ganter, Carl ;
Rummeny, Ernst J. ;
Schwaiger, Markus .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (06) :845-855