Dual amperometric-potentiometric biosensor detection system for monitoring organophosphorus neurotoxins

被引:45
作者
Wang, J
Krause, R
Block, K
Musameh, M
Mulchandani, A
Mulchandani, P
Chen, W
Schöning, MJ
机构
[1] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
[2] Univ Calif Riverside, Dept Chem & Environm Engn, Riverside, CA 92521 USA
[3] Univ Appl Sci, D-82428 Aachen, Julich, Germany
关键词
biosensor; organophosphorous hydrolase; enzyme electrode; pesticides; nerve agents;
D O I
10.1016/S0003-2670(02)00666-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A dual-transducer flow-injection biosensor detection system for monitoring organophosphorus (OP) neurotoxins is described. Such simultaneous use of different physical transducers in connection to the same (organophosphorous hydrolase (OPH)) enzyme enhances the information content and provides discrimination between various subclasses of OP compounds. While the potentiometric biosensor responds favorably to all OP compounds, reflecting the pH changes associated with the OPH activity, the amperometric device displays well-defined signals only towards OP substrates (pesticides) liberating the oxidizable p-nitrophenol product. The potentiometric detection has been accomplished with a silicon-based pH-sensitive electrolyte-insulator-semiconductor (EIS) transducer, operated in the constant-capacitance (ConCap) mode. Both transducers are prepared by a thin-film fabrication technology, and respond rapidly and independently to sudden changes in the level of the corresponding OP compound, with no apparent cross reactivity. Relevant experimental variables were evaluated and optimized. Such development holds great promise for field screening of OP neurotoxins in connection to various defense and environmental scenarios. The multiple-transduction concept could be extended for increasing the information content of other 'class-enzyme' biosensor systems. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:197 / 203
页数:7
相关论文
共 17 条
[1]  
FASSBENDER F, 2002, SENSOR ACTUAT B-CHEM, V68, P128
[2]   AMPEROMETRIC FLOW-THROUGH BIOSENSOR FOR THE DETERMINATION OF PESTICIDES [J].
LAROSA, C ;
PARIENTE, F ;
HERNANDEZ, L ;
LORENZO, E .
ANALYTICA CHIMICA ACTA, 1995, 308 (1-3) :129-136
[3]   Amperometric thick film strip electrodes for monitoring organophosphate nerve agents based on immobilized organophosphorus hydrolase [J].
Mulchandani, A ;
Mulchandani, P ;
Chen, W ;
Wang, J ;
Chen, L .
ANALYTICAL CHEMISTRY, 1999, 71 (11) :2246-2249
[4]   Biosensor for direct determination of organophosphate nerve agents using recombinant Escherichia coli with surface-expressed organophosphorus hydrolase.: 1.: Potentiometric microbial electrode [J].
Mulchandani, A ;
Mulchandani, P ;
Kaneva, I ;
Chen, W .
ANALYTICAL CHEMISTRY, 1998, 70 (19) :4140-4145
[5]   Fiber-optic enzyme biosensor for direct determination of organophosphate nerve agents [J].
Mulchandani, A ;
Pan, ST ;
Chen, W .
BIOTECHNOLOGY PROGRESS, 1999, 15 (01) :130-134
[6]   Biosensors for direct determination of organophosphate pesticides [J].
Mulchandani, A ;
Chen, W ;
Mulchandani, P ;
Wang, J ;
Rogers, KR .
BIOSENSORS & BIOELECTRONICS, 2001, 16 (4-5) :225-230
[7]   Biosensor for direct determination of organophosphate nerve agents. 1. Potentiometric enzyme electrode [J].
Mulchandani, P ;
Mulchandani, A ;
Kaneva, I ;
Chen, W .
BIOSENSORS & BIOELECTRONICS, 1999, 14 (01) :77-85
[8]   ENZYMATIC DETOXIFICATION OF WASTE ORGANO-PHOSPHATE PESTICIDES [J].
MUNNECKE, DM .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1980, 28 (01) :105-111
[9]   Biosensors for chemical and biological agents of defence interest [J].
Paddle, BM .
BIOSENSORS & BIOELECTRONICS, 1996, 11 (11) :1079-1113
[10]  
Poghossian A, 2001, SENSOR MATER, V13, P207