Analog circuit for real-time computation of respiratory mechanical impedance in sleep studies

被引:14
作者
Farre, R [1 ]
Rotger, M [1 ]
Montserrat, JM [1 ]
Navajas, D [1 ]
机构
[1] UNIV BARCELONA,FAC MED,SERV PNEUMOL,HOSP CLIN,E-08036 BARCELONA,SPAIN
关键词
airway resistance; forced oscillation; monitoring; sleep apnea;
D O I
10.1109/10.641343
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this work was to develop a low-cost circuit for real-time analog computation of the respiratory mechanical impedance in sleep studies. The practical performance of the circuit was tested in six patients with obstructive sleep apnea. The impedance signal provided by the analog circuit was compared with the impedance calculated simultaneously with a conventional computerized system. We concluded that the ion-cost analog circuit developed could be a useful tool for facilitating the real-time assessment of airway obstruction in routine sleep studies.
引用
收藏
页码:1156 / 1159
页数:4
相关论文
共 14 条
[1]  
CARSKADON MA, 1994, PRINCIPLES PRACTICE, P943
[2]   TIME-DOMAIN DIGITAL-FILTER TO IMPROVE SIGNAL-TO-NOISE RATIO IN RESPIRATORY IMPEDANCE MEASUREMENTS [J].
FARRE, R ;
ROTGER, M ;
NAVAJAS, D .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (01) :18-24
[3]   Inspiratory dynamic obstruction detected by forced oscillation during CPAP - A model study [J].
Farre, R ;
Peslin, R ;
Rotger, M ;
Navajas, D .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1997, 155 (03) :952-956
[4]   A system to generate simultaneous forced oscillation and continuous positive airway pressure [J].
Farre, R ;
Rotger, M ;
Montserrat, JM ;
Navajas, D .
EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (06) :1349-1353
[5]   SERVOCONTROLLED GENERATOR TO MEASURE RESPIRATORY IMPEDANCE FROM 0.25 TO 26 HZ IN VENTILATED PATIENTS AT DIFFERENT PEEP LEVELS [J].
FARRE, R ;
FERRER, M ;
ROTGER, M ;
NAVAJAS, D .
EUROPEAN RESPIRATORY JOURNAL, 1995, 8 (07) :1222-1227
[6]   COMPUTATION OF RESPIRATORY IMPEDANCE FROM FORCED SINUSOIDAL OSCILLATIONS DURING BREATHING [J].
HOROWITZ, JG ;
SIEGEL, SD ;
PRIMIANO, FP ;
CHESTER, EH .
COMPUTERS AND BIOMEDICAL RESEARCH, 1983, 16 (06) :499-521
[7]  
KEENAN SA, 1994, SLEEP DISORDERS MED, P79
[8]   A NEW ESTIMATOR TO MINIMIZE THE ERROR DUE TO BREATHING IN THE MEASUREMENT OF RESPIRATORY IMPEDANCE [J].
NAVAJAS, D ;
FARRE, R ;
ROTGER, M ;
PESLIN, R .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (12) :1001-1005
[9]   RESPIRATORY INPUT IMPEDANCE IN ANESTHETIZED PARALYZED PATIENTS [J].
NAVAJAS, D ;
FARRE, R ;
CANET, J ;
ROTGER, M ;
SANCHIS, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1990, 69 (04) :1372-1379
[10]  
NAVAJAS D, 1996, AM J RESP CRIT CARE, V153, pA772