Optimization of dispersive liquid-liquid microextraction combined with gas chromatography for the analysis of nitroaromatic compounds in water

被引:89
作者
Ebrahimzadeh, Homeira [1 ]
Yamini, Yadollah [2 ]
Kamarei, Fahimeh [1 ]
机构
[1] Shahid Beheshti Univ, Dept Chem, GC, Tehran, Iran
[2] Tarbiat Modares Univ, Dept Chem, Tehran, Iran
关键词
Dispersive liquid-liquid microextraction; Nitroaromatic compounds; Plackett-Burman design; Gas chromatography; SOLVENT MICROEXTRACTION; EXPLOSIVES; MONONITROTOLUENES; EXTRACTION;
D O I
10.1016/j.talanta.2009.06.022
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学];
摘要
Dispersive liquid-liquid microextraction (DLLME) coupled with gas chromatography-flame ionization detector (GC-FID) was developed for preconcentration and determination of some nitroaromatic compounds in wastewater samples. The effects of different variables on the extraction efficiency were studied simultaneously using experimental design. The variables of interest in the DLLME process were extraction and disperser solvent volumes, salt effect, sample volume, extraction temperature and extraction time. A Plackett-Burman design was performed for screening of variables in order to determine the significant variables affecting the extraction efficiency. Then, the significant factors were optimized by using a central composite design (CCD) and the response surface equations were derived. The optimum experimental conditions found from this statistical evaluation included: sample volume, 9 mL; extraction solvent (CCl4) volume, 20 mu L; disperser solvent (methanol) volume, 0.75 mL; sodium chloride concentration. 3% (w/v); extraction temperature, 20,C and extraction time, 2 min. Under the optimum conditions, the preconcentration factors were between 202 and 314. Limit of detections (LODs) ranged from 0.09 mu g L-1 (for 2-nitrotoluene) to 0.5 mu g L-1 (for 2,4-dinitrotoluene). Linear dynamic ranges (LDRs) of 0.5-300 and 1-400 mu g L-1 were obtained for mononitrotoluenes (MNTs) and dinitrotoluenes (DNTs), respectively. Performance of the present method was evaluated for extraction and determination of nitroaromatic compounds in wastewater samples in the range of microgram per liter and satisfactory results were obtained (RSDs < 10.1%). (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:1472 / 1477
页数:6
相关论文
共 22 条
[1]
Determination of organophosphorus pesticides in water samples by single drop microextraction and gas chromatography-flame photometric detector [J].
Ahmadi, F ;
Assadi, Y ;
Hosseini, SMRM ;
Rezaee, M .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1101 (1-2) :307-312
[2]
Dispersive liquid-liquid microextraction combined with gas chromatography-flame photometric detection - Very simple, rapid and sensitive method for the determination of organophosphorus pesticides in water [J].
Berijani, Sana ;
Assadi, Yaghoub ;
Anbia, Mansoor ;
Milani Hosseini, Mohammad-Reza ;
Aghaee, Elham .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1123 (01) :1-9
[3]
Anaerobic biodegradation of explosives and related compounds by sulfate-reducing and methanogenic bacteria: A review [J].
Boopathy, R ;
Kulpa, CF ;
Manning, J .
BIORESOURCE TECHNOLOGY, 1998, 63 (01) :81-89
[4]
Homogeneous liquid-liquid extraction of trace amounts of mononitrotoluenes from waste water samples [J].
Ebrahimzadeh, Homeira ;
Yamini, Yadollah ;
Kamarei, Fahimeh ;
Shariati, Shahab .
ANALYTICA CHIMICA ACTA, 2007, 594 (01) :93-100
[5]
Application of headspace solvent microextraction to the analysis of mononitrotoluenes in waste water samples [J].
Ebrahimzadeh, Homeira ;
Yamini, Yadollah ;
Kamarei, Fahimeh ;
Khalili-Zanjani, Mohammadreza .
TALANTA, 2007, 72 (01) :193-198
[6]
Dispersive liquid-liquid microextraction combined with graphite furnace atomic absorption spectrometry - Ultra trace determination of cadmium in water samples [J].
Jahromi, Elham Zeini ;
Bidari, Araz ;
Assadi, Yaghoub ;
Hosseini, Mohammad Reza Milani ;
Jamali, Mohammad Reza .
ANALYTICA CHIMICA ACTA, 2007, 585 (02) :305-311
[7]
Part-per-trillion determination of chlorobenzenes in water using dispersive liquid-liquid microextraction combined gas chromatography-electron capture detection [J].
Kozani, Reyhaneh Rahnama ;
Assadi, Yaghoub ;
Shemirani, Farzaneh ;
Hosseini, Mohammad-Reza Milani ;
Jamali, Mohammad Reza .
TALANTA, 2007, 72 (02) :387-393
[8]
Morgan E., 1991, CHEMOMETRICS EXPT DE
[9]
Myers R. H., 2016, RESPONSE SURFACE MET
[10]
Development of a hollow fibre liquid phase microextraction method to monitor the sonochemical degradation of explosives in water [J].
Psillakis, E ;
Mantzavinos, D ;
Kalogerakis, N .
ANALYTICA CHIMICA ACTA, 2004, 501 (01) :3-10