Closed loop control of sedation for colonoscopy using the Bispectral Index

被引:52
作者
Leslie, K [1 ]
Absalom, A
Kenny, GNC
机构
[1] Univ Glasgow, So Gen Hosp, Dept Anaesthesia, Glasgow G31 2ER, Lanark, Scotland
[2] Royal Melbourne Hosp, Dept Anaesthesia & Pain Management, Melbourne, Vic 3050, Australia
[3] Norfolk & Norwich Hosp, Dept Anaesthesia, Norwich NR1 3SR, Norfolk, England
关键词
anaesthetics; intravenous; propofol; anaesthesia; techniques; closed-loop control; measurement; bispectral index;
D O I
10.1046/j.1365-2044.2002.02572_2.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Sixteen patients undergoing colonoscopy were sedated with propofol using a closed-loop control system guided by the Bispectral Index (BIS). Propofol administration, via a target-controlled infusion, was controlled by a proportional-integral-differential control algorithm. The median (range) propofol target concentration during closed-loop control was 2.3 (1.7-3.6) mug.ml(-1).The performance characteristics of the system were excellent, with a median absolute performance error of 7 (1-15)%. Patients were drowsy yet rousable, with a median (range) BIS set-point of 80 (75-85). No patient became apnoeic, required airway support or became haemodynamically unstable whilst sedated. Eight patients moaned or moved during colonoscopy and four had recall. Median (range) time to full consciousness was 4 (2-20) min after the end of closed-loop control. Patient and surgeon satisfaction were high. We conclude that BIS may be a suitable control variable for closed-loop control of sedation with propofol.
引用
收藏
页码:693 / 697
页数:5
相关论文
共 18 条
[1]  
ABSALOM A, 2002, IN PRESS ANESTHESIOL
[2]  
ALDRETE JA, 1970, ANESTH ANAL CURR RES, V49, P924
[3]  
Bell GD, 2001, ENDOSCOPY, V33, P96
[4]  
CHERNIK DA, 1990, J CLIN PSYCHOPHARM, V10, P244
[5]   Automated anesthesia - Fact or fantasy? [J].
Glass, PSA ;
Rampil, IJ .
ANESTHESIOLOGY, 2001, 95 (01) :1-2
[6]   Development and clinical application of electroencephalographic bispectrum monitoring [J].
Johansen, JW ;
Sebel, PS .
ANESTHESIOLOGY, 2000, 93 (05) :1336-1344
[7]   Complications of endoscopy [J].
Kavic, SM ;
Basson, MD .
AMERICAN JOURNAL OF SURGERY, 2001, 181 (04) :319-332
[8]   Closed-loop control of propofol anaesthesia [J].
Kenny, GNC ;
Mantzaridis, H .
BRITISH JOURNAL OF ANAESTHESIA, 1999, 83 (02) :223-228
[9]   PHARMACOKINETIC MODEL DRIVEN INFUSION OF PROPOFOL IN CHILDREN [J].
MARSH, B ;
WHITE, M ;
MORTON, N ;
KENNY, GNC .
BRITISH JOURNAL OF ANAESTHESIA, 1991, 67 (01) :41-48
[10]   Closed loop control of anaesthesia: an assessment of the bispectral index as the target of control [J].
Morley, A ;
Derrick, J ;
Mainland, P ;
Lee, BB ;
Short, TG .
ANAESTHESIA, 2000, 55 (10) :953-959