Experimental investigation of frictional and viscoelastic properties of intestine for microendoscope application

被引:100
作者
Kim, J. -S.
Sung, I. -H.
Kim, Y. -T.
Kwon, E. -Y.
Kim, D. -E.
Jang, Y. H.
机构
[1] Yonsei Univ, Sch Mech Engn, Seoul 120749, South Korea
[2] Hannam Univ, Dept Mech Engn, Taejon 306791, South Korea
[3] Samsung Electromech Co Ltd, Suwon 443743, Gyunggi Do, South Korea
关键词
biotribology; capsule endoscope; small intestine; stress relaxation; viscoelasticity;
D O I
10.1007/s11249-006-9073-0
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The information on the frictional resistance of a self-propelled robotic capsule endoscope moving inside the body is very important for the design and the performance enhancement of such parameters of the capsule endoscope as power consumption, motion control and positioning accuracy. Based on this motivation, the ultimate goal of this research was to develop an analytical model that can predict the frictional resistance of the capsule endoscope moving inside the living body. In this work, experimental investigations of the fundamental frictional characteristics and the viscoelastic behaviors of the small intestine were performed by using custom-built testers and various capsule dummies. The small intestine of a pig was used for the experiments. Experimental results showed that the average frictional force was 10-50 mN and higher moving speed of the capsule dummy resulted in larger frictional resistance of the capsule. In addition, the friction coefficient did not change significantly with respect to the apparent area of contact between the capsule dummy and the intestine, and also the friction coefficients decreased with an increase in the normal load and varied from 0.08 to 0.2. Such frictional behaviors could be explained by the lubrication characteristics of the intestine surface and typical viscoelastic characteristics of the small intestine material. Also, based on the experimental results, a viscoelasticity model for the stress relaxation of the small intestine could be derived.
引用
收藏
页码:143 / 149
页数:7
相关论文
共 19 条
[1]   Models of skeletal muscle to explain the increase in passive stiffness in desmin knockout muscle [J].
Anderson, J ;
Li, Z ;
Goubel, F .
JOURNAL OF BIOMECHANICS, 2002, 35 (10) :1315-1324
[2]  
[Anonymous], FUNDAMENTALS ANATOMY
[3]   Frictional resistance characteristics of a capsule inside the intestine for microendoscope design [J].
Baek, NK ;
Sung, IH ;
Kim, DE .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2004, 218 (H3) :193-201
[4]   Modeling and experimental validation of the locomotion of endoscopic robots in the colon [J].
Dario, P ;
Ciarletta, P ;
Menciassi, A ;
Kim, B .
INTERNATIONAL JOURNAL OF ROBOTICS RESEARCH, 2004, 23 (4-5) :549-556
[5]  
Findley W.N., 1989, CREEP RELAXATION NON, DOI DOI 10.1115/1.3424077
[6]   Capsule endoscopy: The voyage is fantastic - will it change what we do? [J].
Fleischer, DE .
GASTROINTESTINAL ENDOSCOPY, 2002, 56 (03) :452-456
[7]  
Flugge W., 1975, VISCOELASTICITY
[8]  
Fung Y., 1993, BIOMECHANICS MECH PR
[9]   Robotics for surgery [J].
Howe, RD ;
Matsuoka, Y .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 1999, 1 :211-240
[10]   Wireless capsule endoscopy [J].
Iddan, G ;
Meron, G ;
Glukhovsky, A ;
Swain, P .
NATURE, 2000, 405 (6785) :417-417