Calibration studies on dried aerosols for laser ablation inductively coupled plasma mass spectrometry

被引:101
作者
Gunther, D
Cousin, H
Magyar, B
Leopold, I
机构
[1] ETH ZURICH,INST INORGAN CHEM,CH-8092 ZURICH,SWITZERLAND
[2] INST PLANT BIOCHEM,D-06120 HALLE,GERMANY
关键词
laser ablation; inductively coupled plasma mass spectrometry; calibration; aerosols;
D O I
10.1039/a604531f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simultaneous dried solution aerosol (Mistral nebulizer/aerosol dryer) and laser-induced aerosol introduction system was used to investigate the calibration capabilities of dried solutions for LA-ICP-MS. Gas Row rates for the simultaneous systems were optimized and gave the best results with 0.5 l min(-1) for the laser gas flow rate and 0.51 min(-1) for the solution gas flow rate, at,which a sensitivity of 65-80% for LA compared with 1.051 min(-1) single gas flow LA-ICP-MS was maintained, The optimum temperatures for the Mistral nebulizer were 143 degrees C (heating) and -3 degrees C (cooling). Oxide formation (CeO+/Ce+) under these conditions is less than 0.3%. Introduction of enriched Pb-207 (as a solution) and natural lead (tia LA) allows the optimization of both sample introduction systems separately, Analyses were performed on synthetic polyethylene materials, IAEA Soil-7 and CSB-1 reference standards, The RSDs on two sample pellets with five replicates each were better than 10%. Quantitative analyses for all REEs were based on In as the internal standard and Rh as the reference element, Fractionation effects of the internal standard relative to the REE were not observed.
引用
收藏
页码:165 / 170
页数:6