Aortic aneurysmal disease: Assessment of stent-graft treatment - CT versus conventional angiography

被引:125
作者
Armerding, MD [1 ]
Rubin, GD [1 ]
Beaulieu, CF [1 ]
Slonim, SM [1 ]
Olcott, EW [1 ]
Samuels, SL [1 ]
Jorgensen, MJ [1 ]
Semba, CP [1 ]
Jeffrey, RB [1 ]
Dake, MD [1 ]
机构
[1] Stanford Univ, Med Ctr, Sch Med, Dept Radiol, Stanford, CA 94305 USA
关键词
aneurysm; aortic; CT; angiography; comparative studies; computed tomography (CT); interventional procedures; complications; stents and prostheses;
D O I
10.1148/radiology.215.1.r00ap28138
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To compare computed tomographic (CT) angiography and conventional angiography for determining the success of endoluminal stent-graft treatment of aortic aneurysms. MATERIALS AND METHODS: Forty patients underwent conventional angiography and CT angiography following treatment of aortoiliac aneurysms with endoluminal stent-grafts. Six additional sets of conventional angiographic-CT angiographic examinations were performed in five patients after placement of additional stent-grafts or coil embolization to treat perigraft leakage. Three faculty CT radiologists who were blinded to patient clinical data and outcome independently interpreted the CT angiograms, and three faculty angiographers, who were not involved in the stent-graft deployment, interpreted the conventional angiograms. Images were assessed for the presence of postdeployment complications. A reference standard was developed by experienced radiologists using all available images and clinical data. Sensitivities, specificities, and kappa values were calculated. RESULTS: Perigraft leakage was the most commonly identified complication. Twenty perigraft leaks were detected in the results of 46 examinations. Sensitivities and specificities for detecting perigraft leakage were 63% and 77% for conventional angiography and 92% and 90% for CT angiography, respectively. The kappa value was 0.41 for conventional angiography and 0.81 for CT angiography. CONCLUSION: CT angiography is the preferred method for establishing the presence of perigraft leakage following treatment of aortoiliac aneurysms with stent-grafts.
引用
收藏
页码:138 / 146
页数:9
相关论文
共 27 条
[1]  
Adiseshiah M, 1997, J ENDOVASC SURG, V4, P286, DOI 10.1583/1074-6218(1997)004<0286:EROLAA>2.0.CO
[2]  
2
[3]   Abdominal aortic aneurysms: Preliminary technical and clinical results with transfemoral placement of endovascular self-expanding stent-grafts [J].
Blum, U ;
Langer, M ;
Spillner, G ;
Mialhe, C ;
Beyersdorf, F ;
BuitragoTellez, C ;
Voshage, G ;
Duber, C ;
Schlosser, V ;
Cragg, AH .
RADIOLOGY, 1996, 198 (01) :25-31
[4]  
Blum U, 1997, J ENDOVASC SURG, V4, P137, DOI 10.1583/1074-6218(1997)004<0137:TCGEWA>2.0.CO
[5]  
2
[6]   TRANSLUMINAL PLACEMENT OF ENDOVASCULAR STENT-GRAFTS FOR THE TREATMENT OF DESCENDING THORACIC AORTIC-ANEURYSMS [J].
DAKE, MD ;
MILLER, DC ;
SEMBA, CP ;
MITCHELL, RS ;
WALKER, PJ ;
LIDDELL, RP .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (26) :1729-1734
[7]   Magnetic resonance imaging and MR angiography of endoluminally treated abdominal aortic aneurysms [J].
Engellau, L ;
Larsson, EM ;
Albrechtsson, U ;
Jonung, T ;
Ribbe, E ;
Thorne, J ;
Zdanowski, Z ;
Norgren, L .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1998, 15 (03) :212-219
[8]  
Fleiss J. L, 1981, STAT METHODS RATES P, P212
[9]  
HANLEY JA, 1989, CRIT REV DIAGN IMAG, V29, P307
[10]   Renal artery stenosis: Detection and quantification with spiral CT angiography versus optimized digital subtraction angiography [J].
Kaatee, R ;
Beek, FJA ;
deLange, EE ;
vanLeeuwen, MS ;
Smits, HFM ;
vanderVen, PJG ;
Beutler, JJ ;
Mali, WPTM .
RADIOLOGY, 1997, 205 (01) :121-127