Geometrical accuracy and fusion of multimodal vascular images: A phantom study

被引:21
作者
Boussion, N
Soulez, G
De Guise, JA
Daronat, M
Qin, Z
Cloutier, G
机构
[1] Univ Montreal Hosp, Lab Biorheol & Med Ultrason, Res Ctr, Montreal, PQ H2L 2W5, Canada
[2] Univ Montreal Hosp, Dept Radiol, Montreal, PQ H2L 2W5, Canada
[3] Univ Montreal Hosp, Res Ctr, Lab Imaging & Orthoped, Montreal, PQ H2L 2W5, Canada
关键词
vascular phantom; medical imaging; image fusion; geometrical accuracy; ultrasonography; magnetic resonance imaging; computerized tomography; x-ray angiography;
D O I
10.1118/1.1751124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this work was to compare the geometrical accuracy of x-ray angiography, magnetic resonance imaging (MRI), x-ray computed tomography (XCT), and ultrasound imaging (B-mode and IVUS, or intravascular ultrasound) for measuring the lumen diameters of blood vessels. An image fusion method was also developed to improve these measurements. The images were acquired from a phantom that mimic vessels of known diameters. After acquisition, the multimodal images were coregistered by manual alignment of fiducial markers, and then by maximization of mutual information. The fusion method was performed by means of a fuzzy logic modeling approach followed by a combination process based on a possibilistic theory. The results showed (i) the better geometrical accuracy of XCT and IVUS compared to the other modalities, and (ii) the better accuracy and smaller variability of fused images compared to single modalities, with respect to most diameters investigated. For XCT, the error varied from 0.4% to 5.4%, depending on the vessel diameter that ranged from 0.93 to 6.24 mm. For IVUS, the error ranged from -0.3% to 1.7% but the smallest vessel (0.93 mm) could not be investigated because of the probe size. Compared to others fusion schemes, the XCT-MRI fused images provided the best results for both accuracy (from -1.6% to 0.2% for the three largest vessels) and robustness (mean relative error of 1.9%). To conclude, this work underlined both the usefulness of the multimodality vascular phantom as a validation tool and the utility of image fusion in the vascular context. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:1434 / 1443
页数:10
相关论文
共 27 条
[1]   Improving clinical decision support through case-based data fusion [J].
Azuaje, F ;
Dubitzky, W ;
Black, N ;
Adamson, K .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (10) :1181-1185
[2]   Automatic segmentation of subcortical brain structures in MR images using information fusion [J].
Barra, V ;
Boire, JY .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (07) :549-558
[3]  
Bendib K, 1999, J RADIOL, V80, P1561
[4]   Automatic evaluation of arterial diameter variation from vascular echographic images [J].
Beux, F ;
Carmassi, S ;
Salvetti, MV ;
Ghiadoni, L ;
Huang, Y ;
Taddei, S ;
Salvetti, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (12) :1621-1629
[5]   FUZZY MATHEMATICAL MORPHOLOGIES - A COMPARATIVE-STUDY [J].
BLOCH, I ;
MAITRE, H .
PATTERN RECOGNITION, 1995, 28 (09) :1341-1387
[6]   Information combination operators for data fusion: A comparative review with classification [J].
Bloch, I .
IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS PART A-SYSTEMS AND HUMANS, 1996, 26 (01) :52-67
[7]   Power Doppler imaging: Initial evaluation as a screening examination for carotid artery stenosis [J].
Bluth, EI ;
Sunshine, JH ;
Lyons, JB ;
Beam, CA ;
Troxclair, LA ;
Althans-Kopecky, L ;
Crewson, PE ;
Sullivan, MA ;
Smetherman, DH ;
Heidenreich, PA ;
Neiman, HL ;
Burkhardt, JH .
RADIOLOGY, 2000, 215 (03) :791-800
[8]   DISTANCE TRANSFORMATIONS IN ARBITRARY DIMENSIONS [J].
BORGEFORS, G .
COMPUTER VISION GRAPHICS AND IMAGE PROCESSING, 1984, 27 (03) :321-345
[9]  
Bouchon-Meunier B, 2000, NONLINEAR BIOMEDICAL, V1, P1, DOI [10.1002/9780470545362.ch1, DOI 10.1002/9780470545362.CH1]
[10]   Extraction of epileptogenic foci from PET and SPECT images by fuzzy modeling and data fusion [J].
Boussion, N ;
Cinotti, L ;
Barra, V ;
Ryvlin, P ;
Mauguiere, F .
NEUROIMAGE, 2003, 19 (03) :645-654