Development of an implantable undulation type ventricular assist device for control of organ circulation

被引:6
作者
Yambe, T
Abe, Y
Imachi, K
Shiraishi, Y
Shibata, MI
Yamaguchi, T
Wang, QT
Duan, XD
Liu, HJ
Yoshizawa, M
Tanaka, A
Matsuki, H
Sato, F
Haga, YI
Esashi, M
Tabayashi, K
Mitamura, Y
Sasada, H
Umezu, M
Matsuda, T
Nitta, SI
机构
[1] Univ Tokyo, Grad Sch Med, Tokyo, Japan
[2] Tohoku Univ, Informat Synergy Ctr, Tohoku, Japan
[3] Tohoku Univ, Inst Dev Aging & Canc, Dept Med Engn & Cardiol, Tohoku, Japan
[4] Tohoku Univ, Grad Sch Engn, Tohoku, Japan
[5] Tohoku Univ, Grad Sch Med, Tohoku, Japan
[6] Hokkaido Univ, Grad Sch Engn, Sapporo, Hokkaido 060, Japan
[7] Tohoku Univ, Grad Sch Agr, Tohoku, Japan
[8] Waseda Univ, Grad Sch Sci & Engn, Waseda, Japan
[9] Kyushu Univ, Grad Sch Med, Kyushu, Japan
关键词
undulation pump; ventricular assist device; organ circulation; multiple organ failure; laser doppler flowmeter;
D O I
10.1111/j.1525-1594.2004.07382.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
It is well known that a rotary blood pump (RP) is effective as a small ventricular assist device (VAD). It might be still more effective if pulsation was available. The undulation pump (UP), which is a type of small RP, can also produce pulsation. In Japan, a development project for an implantable type UP ventricular assist device (UPVAD) is now advanced. Six universities and some companies together have been in charge of the development project for 5 years. In this study, the influence which the UP under development has on circulation in internal organs was investigated. Goats with the same weight as an average Asian person were used for the experiment. The left chest cavity was opened after resection of the fourth rib and the heart was approached. A cannula was inserted in the left ventricle from the apex. An outflow cannula was inserted into the left descending aorta. Heart muscle was excised using a newly developed puncher. The UPVAD was implanted using a left-heart bypass system. The myocardial blood flow, carotid arterial blood flow, and the kidney blood flow were recorded together with an electrocardiogram, blood pressure, and the flow rate. In these animal experiments, the blood circulation dynamic state was stabilized and sufficient support of the left heart was observed. Myocardial blood flow, carotid arterial flow, and a kidney blood flow increase resulting from UPVAD support was observed. Often the problem of multiple organ failure is important at the time of clinical application of a ventricular assist device. Assisting circulation to internal organs is important for prevention of multiple organ failure. It was concluded that the UPVAD might be useful for prevention of multiple organ failure.
引用
收藏
页码:940 / 944
页数:5
相关论文
共 20 条
[1]
Advance in animal experiments with the undulation pump total artificial heart: 50 and 54 day survival periods with 1/R control [J].
Abe, Y ;
Chinzei, T ;
Isoyama, T ;
Kobayashi, S ;
Ono, T ;
Saito, I ;
Iwasaki, K ;
Ishimaru, M ;
Baba, A ;
Kouno, A ;
Ozeki, T ;
Tohyama, T ;
Imachi, K .
ASAIO JOURNAL, 2003, 49 (03) :325-332
[2]
Continuously maintaining positive flow avoids endocardial suction of a rotary blood pump with left ventricular bypass [J].
Akimoto, T ;
Yamazaki, K ;
Litwak, P ;
Litwak, KN ;
Tagusari, O ;
Mori, T ;
Antaki, JF ;
Kameneva, MV ;
Watach, MJ ;
Umezu, M ;
Tomioka, J ;
Kormos, RL ;
Koyanagi, H ;
Griffith, BP .
ARTIFICIAL ORGANS, 2000, 24 (08) :606-610
[3]
Mesana T, 1991, ASAIO Trans, V37, pM166
[4]
Study of left ventricular bypass using Wankel type semipulsatile blood pump [J].
Mitsui, N ;
Fukunaga, S ;
Sueda, T ;
Matsuura, Y ;
Havlik, P ;
Trinkl, J ;
Demunck, JL ;
Mesana, T ;
Monties, JR .
ARTIFICIAL ORGANS, 1998, 22 (05) :419-425
[5]
Monties J R, 1990, ASAIO Trans, V36, pM258
[6]
Cora valveless pulsatile rotary pump: New design and control [J].
Monties, JRE ;
Trinkl, J ;
Mesana, T ;
Havlik, PJ ;
Demunck, JLY .
ANNALS OF THORACIC SURGERY, 1996, 61 (01) :463-468
[7]
Can we develop a nonpulsatile permanent rotary blood pump? Yes, we can. [J].
Nose, Y ;
Kawahito, K ;
Nakazawa, T .
ARTIFICIAL ORGANS, 1996, 20 (06) :467-474
[8]
Development of an implantable high-energy and compact battery system for artificial heart [J].
Okamoto, E ;
Inoue, T ;
Watanabe, K ;
Hashimoto, T ;
Iwazawa, E ;
Abe, Y ;
Chinzei, T ;
Isoyama, T ;
Kobayashi, S ;
Saito, I ;
Sato, F ;
Matsuki, H ;
Imachi, K ;
Mitamura, Y .
ARTIFICIAL ORGANS, 2003, 27 (02) :184-188
[9]
Functions for detecting malposition of transcutaneous energy transmission coils [J].
Ozeki, T ;
Chinzei, T ;
Abe, Y ;
Saito, I ;
Isoyama, T ;
Mochizuki, S ;
Ishimaru, M ;
Takiura, K ;
Baba, A ;
Toyama, T ;
Imachi, K .
ASAIO JOURNAL, 2003, 49 (04) :469-474
[10]
TECHNICAL REQUIREMENTS AND LIMITATIONS OF MINIATURIZED AXIAL-FLOW PUMPS FOR CIRCULATORY SUPPORT [J].
REUL, H .
CARDIOLOGY, 1994, 84 (03) :187-193