Investigation of microwave assisted drying of samples and evaporation of aqueous solutions in trace element analysis

被引:17
作者
Maichin, B
Kettisch, P
Knapp, G
机构
[1] Graz Tech Univ, Inst Analyt Chem Micro & Radiochem, A-8010 Graz, Austria
[2] Anton Paar GmbH, A-8054 Graz, Austria
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 2000年 / 366卷 / 01期
关键词
D O I
10.1007/s002160050006
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Investigations of microwave assisted drying of sample materials and microwave assisted evaporation of aqueous sample solutions and acidic digestion residues were accomplished by means of special rotors for the microwave digestion system MULTIWAVE. To check the results obtained by microwave assisted drying, the samples were also conventionally dried at 105 degrees C in an oven. The following samples have been dried: 10 g each of meat, fish, apple, cucumber, potato, mustard, yogurt, clay and marl; 1 g each of certified reference material TORT 2 (lobster hepatopancreas), BCR-278 (mussel tissue) and BCR 422 (cod muscle); 500 g garden mould. Microwave assisted drying takes 40 min for organic samples and 30 min for inorganic material. Important is a slow increase of microwave power during the first 20 min. The results agree well with conventional drying at 105 degrees C. Losses of As, Se and Hg have been investigated for 3 CRMs. Only Se shows losses in the range of 20%. Losses of As, Be, Cd, Co, Cr, Cu, Fe, Ng, Li, Mg, Mn, Mo, Ni, Pb, Sb, Se, Sr, Ti, Tl, V and Zn after evaporation of aqueous samples and acidic solutions after wet digestion, respectively, have been investigated. 50 mL aqueous solution was evaporated almost to dryness within 25 min. The recovery of Hg is 40-50%, of Se 90-95% and of the other elements 97-102%. 0.2 g each of TORT 2, BCR 278 and BCR 422 have been digested with 4 mL nitric acid and 1 mL hydrochloric acid by means of the microwave digestion system MULTIWAVE. The digestion residue was evaporated almost to dryness and dissolved again in 10 mt diluted nitric acid. In this case no element losses have been observed. The measured concentration of As, Cd, Cu, Fe, Mn, Hg, Pb, Mo, Ni, Se, Sr, V and Zn agree very well with the certified values. An important prerequisite for good recoveries is not to evaporate the solutions to complete dryness.
引用
收藏
页码:26 / 29
页数:4
相关论文
共 13 条
[1]  
ALFASSI ZB, 1994, DETERMINATION TRACE, P13
[3]  
FEIG H, 1995, GIT SEP, P815
[4]   USE OF A MICROWAVE OVEN IN IN-PROCESS CONTROL [J].
HESEK, JA ;
WILSON, RC .
ANALYTICAL CHEMISTRY, 1974, 46 (08) :1160-1160
[5]   MICROWAVE DRYING OF BIOLOGICAL TISSUES FOR TRACE-ELEMENT DETERMINATION [J].
KOH, TS .
ANALYTICAL CHEMISTRY, 1980, 52 (12) :1978-1979
[6]   APPLICATION OF A MICROWAVE-OVEN FOR DRYING AND NITRIC-ACID EXTRACTION OF MERCURY AND SELENIUM FROM FISH TISSUE [J].
LAMLEUNG, SY ;
CHENG, VKW ;
LAM, YW .
ANALYST, 1991, 116 (09) :957-959
[7]  
Muller I, 1998, ATOM SPECTROSC, V19, P45
[8]   Elemental analysis using microwave digestion and an environmental evaporation chamber [J].
Rhoades, CB ;
Levine, KE ;
Salido, A ;
Jones, BT .
APPLIED SPECTROSCOPY, 1998, 52 (02) :200-204
[9]   APPLICATION OF THE MICROWAVE-OVEN TO THE PRETREATMENT OF MACROSAMPLES IN ENVIRONMENTAL-RADIOACTIVITY MONITORING [J].
ROIG, M ;
RIBERA, MM ;
RAURET, G .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1995, 190 (01) :59-69
[10]  
SENGUTTA U, 1998, LABO DEZ, P30