Electrical stimulation of isolated retina with microwire glass electrodes

被引:32
作者
Johnson, L [1 ]
Perkins, FK
O'Hearn, T
Skeath, P
Merritt, C
Frieble, J
Sadda, S
Humayun, M
Scribner, D
机构
[1] USN, Res Lab, Washington, DC USA
[2] Johns Hopkins Univ, Wilmer Eye Inst, Baltimore, MD 21218 USA
[3] Univ So Calif, Doheny Eye Inst, Los Angeles, CA 90033 USA
关键词
retinal prosthesis; isolated retina; stimulating electrodes; blindness; channel glass;
D O I
10.1016/j.jneumeth.2004.02.035
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The development of high-resolution retina] prostheses fabricated from silicon wafers presents an interesting problem: how to electrically bridge the space between the flat silicon wafer and the curved retinal surface. One potential "bridge" is a microwire glass electrode. In this paper we present our results in evaluating microwire glass electrodes. We stimulated isolated rabbit retina (n = 5) with a 0.0256 cm(2) microwire electrode. The current and pulse duration were varied from 498 to 1660 muA and 0.1 to 3 ms, respectively. We found that short pulses produced more spikes per coulomb and longer pulses produced more spikes per milliamp. The optimal pulse duration range of 0.7-1 ms was identified as a compromise between the advantages of short and long pulses. Stimulation of isolated rabbit retina with microwire glass results in consistent neuronal spike formation at safe charge density, 20.7 +/- 14.3 muC/cm(2). We also examined the response of retinas (n = 6) to stimulation with a smaller microwire electrode, 0.0002 cm(2). We found that less current was required (15 muA versus 756 muA) for a I ms pulse, but at the expense of greater charge density (75 muC/cm(2) versus 29.5 muC/cm(2)). Nonetheless, a 128-fold reduction in area resulted in only a 2.7-fold increase in charge density required for a 1 ms pulse duration. The results presented here indicate that microwire glass can be used as a neural stimulating electrode to bridge the gap between flat microelectronic stimulator chips and curved neuronal tissue. (C) 2004 Published by Elsevier B.V.
引用
收藏
页码:265 / 273
页数:9
相关论文
共 33 条
[1]   INVITRO RETINA AS AN EXPERIMENTAL-MODEL OF THE CENTRAL NERVOUS-SYSTEM [J].
AMES, A ;
NESBETT, FB .
JOURNAL OF NEUROCHEMISTRY, 1981, 37 (04) :867-877
[2]   THE STRENGTH DURATION RELATIONSHIP FOR EXCITATION OF MYELINATED NERVE - COMPUTED DEPENDENCE ON MEMBRANE PARAMETERS [J].
BOSTOCK, H .
JOURNAL OF PHYSIOLOGY-LONDON, 1983, 341 (AUG) :59-74
[3]   Subretinal electrical stimulation of the rabbit retina [J].
Chow, AY ;
Chow, VY .
NEUROSCIENCE LETTERS, 1997, 225 (01) :13-16
[4]   Implantation of silicon chip microphotodiode arrays into the cat subretinal space [J].
Chow, AY ;
Pardue, MT ;
Chow, VY ;
Peyman, GA ;
Liang, CP ;
Perlman, JI ;
Peachey, NS .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2001, 9 (01) :86-95
[5]   LOGISTIC PATTERNING OF BRAIN ACTIVITY [J].
DESHMUKH, VD .
CLINICAL ELECTROENCEPHALOGRAPHY, 1989, 20 (04) :266-272
[6]   Learning retina implants with epiretinal contacts [J].
Eckmiller, R .
OPHTHALMIC RESEARCH, 1997, 29 (05) :281-289
[7]   The effect of stimulus pulse duration on selectivity of neural stimulation [J].
Grill, WM ;
Mortimer, JT .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (02) :161-166
[8]   Multi-electrode stimulation and recording in the isolated retina [J].
Grumet, AE ;
Wyatt, JL ;
Rizzo, JF .
JOURNAL OF NEUROSCIENCE METHODS, 2000, 101 (01) :31-42
[9]   BIPOLAR SURFACE ELECTRICAL-STIMULATION OF THE VERTEBRATE RETINA [J].
HUMAYUN, M ;
PROPST, R ;
DEJUAN, E ;
MCCORMICK, K ;
HICKINGBOTHAM, D .
ARCHIVES OF OPHTHALMOLOGY, 1994, 112 (01) :110-116
[10]  
Humayun M S, 2001, Trans Am Ophthalmol Soc, V99, P271