Immobilized parathion hydrolase: An amperometric sensor for parathion

被引:65
作者
Sacks, V
Eshkenazi, I
Neufeld, T
Dosoretz, C
Rishpon, J [1 ]
机构
[1] Tel Aviv Univ, Dept Mol Microbiol & Biotechnol, IL-69978 Ramat Aviv, Israel
[2] Acad Coll, Sch Sci & Technol, Dept Biotechnol & Environm Sci, Tel Hai, Upper Galilee, Israel
[3] Migal Galilee Technol Ctr, Kiryat Shmona, Israel
关键词
D O I
10.1021/ac9911488
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric enzyme biosensor for the direct measurement of parathion was developed. The biosensor is based on parathion hydrolase from Pseudomonas sp. isolated from contaminated soil. The enzyme, which was immobilized on a carbon electrode, catalyzes the hydrolysis of parathion to form p-nitrophenol, which was detected by its anodic oxidation. The enzymatic and electrochemical reactions were examined and optimized. Screen-printed electrodes and a microflow injection system provide the means to significantly reduce the volume of the detected samples. Pulsed techniques further increased the sensitivity of the measurement. The current signal was linearly related to the parathion concentration, and the detection limit was less than 1 ng/mL. The biosensor is rapid as well and can be used outdoors and indoors by a nonqualified person.
引用
收藏
页码:2055 / 2058
页数:4
相关论文
共 22 条
[1]  
*AM C GOV IND HYG, 1994, THRESHH LIM VAL CHEM
[2]   PHOSPHOTRIESTERASES OF FLAVOBACTERIUM SP [J].
BROWN, KA .
SOIL BIOLOGY & BIOCHEMISTRY, 1980, 12 (02) :105-112
[3]   Entrapment of parathion hydrolase from Pseudomonas spp in sol-gel glass [J].
Dosoretz, C ;
Armon, R ;
Starosvetzky, J ;
Rothschild, N .
JOURNAL OF SOL-GEL SCIENCE AND TECHNOLOGY, 1996, 7 (1-2) :7-11
[4]  
DUMAS DP, 1989, J BIOL CHEM, V264, P19659
[5]   INACTIVATION OF ORGANOPHOSPHORUS NERVE AGENTS BY THE PHOSPHOTRIESTERASE FROM PSEUDOMONAS-DIMINUTA [J].
DUMAS, DP ;
DURST, HD ;
LANDIS, WG ;
RAUSHEL, FM ;
WILD, JR .
ARCHIVES OF BIOCHEMISTRY AND BIOPHYSICS, 1990, 277 (01) :155-159
[6]  
DURAND P, 1984, J ENVIRON PATHOL TOX, V5, P51
[7]   Increased paraoxon detection with solvents using acetylcholinesterase inactivation measured with a choline oxidase biosensor [J].
Fennouh, S ;
Casimiri, V ;
Burstein, C .
BIOSENSORS & BIOELECTRONICS, 1997, 12 (02) :97-104
[8]  
Hayes WJ, 1990, HDB PESTICIDE TOXICO, V3
[9]   Biosensors for environmental monitoring [J].
Karube, I ;
Yano, K ;
Sasaki, S ;
Nomura, Y ;
Ikebukuro, K .
ENZYME ENGINEERING XIV, 1998, 864 :23-36
[10]   VALIDATION OF AN AUTOMATED PRECOLUMN EXCHANGE SYSTEM (PROSPEKT) COUPLED TO LIQUID-CHROMATOGRAPHY WITH DIODE-ARRAY DETECTION - APPLICATION TO THE DETERMINATION OF PESTICIDES IN NATURAL-WATERS [J].
LACORTE, S ;
BARCELO, D .
ANALYTICA CHIMICA ACTA, 1994, 296 (03) :223-234