Characterization in vitro and in vivo of the oxygen dependence of an enzyme/polymer biosensor for monitoring brain glucose

被引:89
作者
Dixon, BM
Lowry, JP
O'Neill, RD [1 ]
机构
[1] Natl Univ Ireland Univ Coll Dublin, Dept Chem, Dublin 4, Ireland
[2] Natl Univ Ireland, Dept Chem, Maynooth, Kildare, Ireland
关键词
poly(o-phenylenediamine); enzyme modified electrode; amperometry; voltammetry in vivo; glucose oxidase; electropolymerization;
D O I
10.1016/S0165-0270(02)00170-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The oxygen dependence of a first generation amperometric biosensor was investigated in vitro and in vivo by monitoring its glucose response as a function of solution PO2. The biosensor was a glucose oxidase (GOx) modified poly(o-phenylenediamine) coated Pt cylinder electrode (Pt/PPD/GOx) that has been designed for neurochemical analysis in vivo. Two types of oxygen probes were used: a self-calibrating commercial macroelectrode in vitro; and a carbon paste microelectrode in vivo. Calibrations in vitro showed that oxygen interference in the operation of Pt/PPD/GOx electrodes was minimal for concentrations of glucose (similar to0.5 mM) and oxygen (similar to50 muM) found in brain ECF. This observation was confirmed by simultaneous monitoring in vivo of brain glucose and oxygen in the awake rat. However, at levels of glucose normally found in peripheral tissues (similar to5 mM), the oxygen dependence was severe. We conclude that the oxygen sensitivity of Pt/PPD/GOx biosensors does not preclude their reliable use in media containing low glucose levels, such as brain ECF. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:135 / 142
页数:8
相关论文
共 52 条
[1]   ELECTROCHEMICAL IMMOBILIZATION OF ENZYMES .2. GLUCOSE-OXIDASE IMMOBILIZED IN POLY-N-METHYLPYRROLE [J].
BARTLETT, PN ;
WHITAKER, RG .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1987, 224 (1-2) :37-48
[2]   A MICROELECTROCHEMICAL ENZYME TRANSISTOR RESPONSIVE TO GLUCOSE [J].
BARTLETT, PN ;
BIRKIN, PR .
ANALYTICAL CHEMISTRY, 1994, 66 (09) :1552-1559
[3]   ELECTROCHEMICAL STUDY OF D-GLUCOSE OXIDASE AUTO-INACTIVATION [J].
BOURDILLON, C ;
THOMAS, V ;
THOMAS, D .
ENZYME AND MICROBIAL TECHNOLOGY, 1982, 4 (03) :175-180
[4]  
Braguglia CM, 1998, CHEM BIOCHEM ENG Q, V12, P183
[5]  
Clark L C Jr, 1965, Ala J Med Sci, V2, P353
[6]   ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY [J].
CLARK, LC ;
LYONS, C .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) :29-&
[7]   HOMOGENEOUS FERROCENE-MEDIATED AMPEROMETRIC IMMUNOASSAY [J].
DIGLERIA, K ;
HILL, HAO ;
MCNEIL, CJ ;
GREEN, MJ .
ANALYTICAL CHEMISTRY, 1986, 58 (06) :1203-1205
[8]  
Dixon BM, 2001, MONITORING MOLECULES IN NEUROSCIENCE, P53
[9]   GLUCOSE EXPORT FROM THE BRAIN IN MAN - EVIDENCE FOR A ROLE FOR ASTROCYTIC GLYCOGEN AS A RESERVOIR OF GLUCOSE FOR NEURAL METABOLISM [J].
EYRE, JA ;
STUART, AG ;
FORSYTH, RJ ;
HEAVISIDE, D ;
BARTLETT, K .
BRAIN RESEARCH, 1994, 635 (1-2) :349-352
[10]   Extracellular glucose turnover in the striatum of unanaesthetized rats measured by quantitative microdialysis [J].
Fray, AE ;
Boutelle, M ;
Fillenz, M .
JOURNAL OF PHYSIOLOGY-LONDON, 1997, 504 (03) :721-726