A prospective randomised multicentre trial comparing 10 Fr Teflon Tannenbaum stents with 10 Fr polyethylene Cotton-Leung stents in patients with malignant common duct strictures

被引:59
作者
England, RE
Martin, DF
Morris, J
Sheridan, MB
Frost, R
Freeman, A
Lawrie, B
Deakin, M
Fraser, I
Smith, K
机构
[1] S Manchester Univ Hosp NHS Trust, Dept Radiol, Manchester M20 2LR, Lancs, England
[2] S Manchester Univ Hosp NHS Trust, Dept Med Stat, Manchester, Lancs, England
[3] St James Hosp, Dept Radiol, Leeds LS9 7TF, W Yorkshire, England
[4] Salisbury Hosp NHS Trust, Dept Radiol, Salisbury, Wilts, England
[5] Addenbrookes Hosp, Dept Radiol, Cambridge CB2 2QQ, England
[6] Univ Wales Hosp, Dept Radiol, Cardiff CF4 4XW, S Glam, Wales
[7] N Staffordshire NHS Trust, Dept Surg, Stoke On Trent, Staffs, England
[8] Walsgrave Gen Hosp, Dept Surg, Coventry CV2 2DY, W Midlands, England
[9] SMS Crowthorne, Crowthorne, Berks, England
关键词
biliary prosthesis; patency; bile duct malignancy;
D O I
10.1136/gut.46.3.395
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background-Stent blockage is a multifactorial process in which stent design and materials, bacteria, proteins, and bile viscosity play a role. Aims-To compare the patency of the 10 Fr Teflon Tannenbaum (TT) stent to that of the 10 Fr Cotton-Leung (CL) polyethylene stent with sideholes, in patients with malignant obstructive jaundice. Methods-Patients were recruited to this prospective multicentre randomised study if they had a newly diagnosed malignant bile duct stricture below the hilum of the liver suitable for stenting with a 10 Fr stent. Data were collected and monitored by a professional monitoring company. Primary patency was the interval between stent placement and first exchange or death without recurrent jaundice. Results-134 consecutive patients were recruited between November 1994 and June 1997; 65 were randomised to the TT stent and 69 to the CL stent. Median patency and 95% confidence intervals were 181 (59, 303) days for the TT stent and 133 (92, 174) days for the CL stent, with no significant difference between the two stents (p=0.49). Median survival and 95% confidence intervals were 115 (71, 159) days for the TT stent and 151 (112, 190) days for the CL stent, with no significant difference between the two stents (p=0.765). Conclusion-Neither Teflon as a stent material nor the Tannenbaum design prolong the patency of plastic stents.
引用
收藏
页码:395 / 400
页数:6
相关论文
共 14 条
  • [1] BRUCKNER M, 1990, FORTSCHR GASTROENTER, V20, P17
  • [2] COENE LO, 1990, ENDOSCOPIC BILIARY S, P135
  • [3] CLOGGING OF BILIARY ENDOPROSTHESES - A NEW PERSPECTIVE
    COENE, PPLO
    GROEN, AK
    CHENG, J
    OUT, MMJ
    TYTGAT, GNJ
    HUIBREGTSE, K
    [J]. GUT, 1990, 31 (08) : 913 - 917
  • [4] THE ROLE OF BACTERIA IN THE BLOCKAGE OF BILIARY STENTS
    LEUNG, JWC
    LING, TKW
    VALLANCEOWEN, J
    [J]. GASTROINTESTINAL ENDOSCOPY, 1988, 34 (01) : 19 - 22
  • [5] ENDOSCOPIC BILIARY PROSTHESES - A COMPARISON OF MATERIALS
    LEUNG, JWC
    DELFAVERO, G
    COTTON, PB
    [J]. GASTROINTESTINAL ENDOSCOPY, 1985, 31 (02) : 93 - 95
  • [6] MARTIN DF, 1998, CLIN PRACTICE ERCP, P53
  • [7] EXPERIMENTAL-STUDY OF BILIARY ENDOPROSTHESIS EFFICIENCY
    REY, JF
    MAUPETIT, P
    GREFF, M
    [J]. ENDOSCOPY, 1985, 17 (04) : 145 - 148
  • [8] PROLONGED PATENCY WITH A NEW-DESIGN TEFLON BILIARY PROSTHESIS
    SEITZ, U
    VADEYAR, H
    SOEHENDRA, N
    [J]. ENDOSCOPY, 1994, 26 (05) : 478 - 482
  • [9] OPTIMAL PALLIATION OF MALIGNANT BILE-DUCT OBSTRUCTION - EXPERIENCE WITH ENDOSCOPIC-12 FRENCH PROSTHESES
    SIEGEL, JH
    PULLANO, W
    KODSI, B
    COOPERMAN, A
    RAMSEY, W
    [J]. ENDOSCOPY, 1988, 20 (04) : 137 - 141
  • [10] BILIARY STENT BLOCKAGE WITH BACTERIAL BIOFILM - A LIGHT AND ELECTRON-MICROSCOPY STUDY
    SPEER, AG
    COTTON, PB
    RODE, J
    SEDDON, AM
    NEAL, CR
    HOLTON, J
    COSTERTON, JW
    [J]. ANNALS OF INTERNAL MEDICINE, 1988, 108 (04) : 546 - 553