Retention time reproducibility in comprehensive two-dimensional gas chromatography using cryogenic modulation - An intralaboratory study

被引:50
作者
Shellie, RA [1 ]
Xie, LL [1 ]
Marriott, PJ [1 ]
机构
[1] RMIT Univ, Dept Appl Chem, Australian Ctr Res Separat Sci, Melbourne, Vic 3001, Australia
关键词
gas chromatography; comprehensive two-dimensional; retention time; cryogenic modulation; reproducibility;
D O I
10.1016/S0021-9673(02)00961-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A survey was conducted to determine the reproducibility of retention times in both the first (D1) and second dimension (D2) axes of the two-dimensional separation space, in the comprehensive two-dimensional gas chromatographic analysis of an essential oil sample using cryogenic modulation. The retention times in the two dimensions for a number of individual components comprising hydrocarbon, alcohol, ester and ketone chemical classes in a Melaleuca alternifolia essential oil were recorded from replicate analyses using four separate column sets and two identical gas chromatographs. Run-to-run, day-to-day, instrument-to-instrument, and column set-to-column set reproducibility were demonstrated from the experimental design. A total of 60 GCxGC analyses were conducted. The longitudinally modulated cryogenic system produced reproducible modulation start times and consistent modulation phase profiles for individual components in all experiments, and retention time variations in both dimensions were negligible. The average run-to-run reproducibility of 43 components for six replicate injections was found to be 0.12% RSD in the first dimension, and 0.74% RSD in the second dimension. Day-to-day reproducibility showed statistically "significant" difference (F-test), but this was partly ascribable to the excellent within-day reproducibility that led to apparent day-to-day differences. Confidence in absolute retention times (hence component positions) in the two-dimensional separation space is critical to component identification. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:161 / 170
页数:10
相关论文
共 28 条
[1]   Moving cryogenic modulator for the comprehensive two-dimensional gas chromatography (GC x GC) of surface water contaminants [J].
Beens, J ;
Dallüge, J ;
Adahchour, M ;
Vreuls, RJJ ;
Brinkman, UAT .
JOURNAL OF MICROCOLUMN SEPARATIONS, 2001, 13 (03) :134-140
[2]   Simple, non-moving modulation interface for comprehensive two-dimensional gas chromatography [J].
Beens, J ;
Adahchour, M ;
Vreuls, RJJ ;
van Altena, K ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 2001, 919 (01) :127-132
[3]   Comprehensive two dimensional high-speed gas chromatography with chemometric analysis [J].
Bruckner, CA ;
Prazen, BJ ;
Synovec, RE .
ANALYTICAL CHEMISTRY, 1998, 70 (14) :2796-2804
[4]  
Davis JM, 2000, HRC-J HIGH RES CHROM, V23, P235, DOI 10.1002/1521-4168(20000301)23:3<235::AID-JHRC235>3.0.CO
[5]  
2-I
[6]   Characterisation of fatty acids in biological oil samples using comprehensive multidimensional gas chromatography [J].
de Geus, HJ ;
Aidos, I ;
de Boer, J ;
Luten, JB ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 2001, 910 (01) :95-103
[7]   Shape selectivity: A key factor in comprehensive two-dimensional gas chromatographic analysis of toxic PCBs [J].
Haglund, P ;
Harju, M ;
Ong, R ;
Marriott, P .
JOURNAL OF MICROCOLUMN SEPARATIONS, 2001, 13 (07) :306-311
[8]  
Kinghorn RM, 2000, HRC-J HIGH RES CHROM, V23, P245
[9]  
Kinghorn RM, 1998, HRC-J HIGH RES CHROM, V21, P620
[10]   A model of peak amplitude enhancement in orthogonal two-dimensional gas chromatography [J].
Lee, AL ;
Bartle, KD ;
Lewis, AC .
ANALYTICAL CHEMISTRY, 2001, 73 (06) :1330-1335