Improvement of abundance sensitivity in a quadrupole-based ICP-MS instrument with a hexapole collision cell

被引:43
作者
Boulyga, SF
Becker, JS [1 ]
机构
[1] Res Ctr Julich, Cent Dept Analyt Chem, D-52425 Julich, Germany
[2] Radiat Phys & Chem Problems Inst, Minsk 220109, BELARUS
关键词
D O I
10.1039/b203086c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High isotopic abundance sensitivity is desirable for the measurement of low-abundance isotopes in the presence of neighboring high-abundance isotopes. This is of interest when analyzing ultra low concentrations of trace elements or isotopes in the presence of elements or isotopes with much higher concentrations (e. g., radionuclides included in a matrix of stable isotopes). In this study, a quadrupole-based inductively coupled plasma mass spectrometer with a hexapole collision cell (ICP-CC-QMS), Platform ICP (Micromass UK), was used to study the improvement of abundance sensitivity using uranium and lutetium as examples. The median kinetic energy of U-238(+) ions was about 1.2 eV and 0.9 eV when helium was introduced as a collision gas at flow rates of 5 ml min(-1) and 10 ml min(-1), respectively. In the latter case, the proportion of ions with relatively high energy decreased significantly, which improved both the transmission of ions through the hexapole collision cell and mass selection by the quadrupole mass spectrometer. When introducing helium into the collision cell a reduction in peak tail of high-abundance isotopes by up to three orders of magnitude, depending on the mass analyzed, was observed. The abundance sensitivity for U-236/U-238 isotope ratio was improved from 2.3x10(-5) to 6.3x10(-8). Abundance sensitivities as low as 3.5x10(-8), 6.6x10(-8) and 2.0x10(-8) were achieved at masses m-2.5 u, m-1.5 u and m+2.5 u (m=174.94 u for Lu-175), respectively. Reducing the kinetic energy also improved the mass resolution by about 10%. However, increasing the mass resolution of the quadrupole instrument had a significantly lower influence on abundance sensitivity than reducing the ion energy by collision with helium atoms. The present study of the effect of the pressurized ion guide on abundance sensitivity was limited to some extent because of the relatively high background count rate of the Daly-type ion detector in the ICP-CC-QMS system. An ion detector possessing a low background would be desirable for further studies. Improved abundance sensitivity can be useful in analyzing long-lived radionuclides like U-236 and I-129 in the presence of high-abundance stable isotopes like U-238 and I-127.
引用
收藏
页码:1202 / 1206
页数:5
相关论文
共 28 条
  • [1] [Anonymous], INTRO MASS SPECTROME
  • [2] [Anonymous], ENCY ANAL CHEM
  • [3] Kinetic energy dependence of ion-molecule reactions: Guided ion beams and threshold measurements
    Armentrout, PB
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2000, 200 (1-3) : 219 - 241
  • [4] Inorganic mass spectrometric methods for trace, ultratrace, isotope, and surface analysis
    Becker, JS
    Dietze, HJ
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2000, 197 : 1 - 35
  • [5] BASIC FINDINGS ON THE USE OF GAS TARGETS FOR THE IMPROVEMENT OF THE SENSITIVITY OF MASS-SPECTROMETRIC METHODS
    BECKER, S
    DIETZE, HJ
    [J]. ISOTOPENPRAXIS, 1983, 19 (04): : 105 - 112
  • [6] Peak shape for a quadrupole mass spectrometer:: comparison of computer simulation and experiment
    Blaum, K
    Geppert, C
    Müller, P
    Nörtershäuser, W
    Wendt, K
    Bushaw, BA
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2000, 202 (1-3) : 81 - 89
  • [7] Properties and performance of a quadrupole mass filter used for resonance ionization mass spectrometry
    Blaum, K
    Geppert, C
    Müller, P
    Nörtershäuser, W
    Otten, EW
    Schmitt, A
    Trautmann, N
    Wendt, K
    Bushaw, BA
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1998, 181 : 67 - 87
  • [8] EFFECT OF CHARGE-EXCHANGE REACTIONS ON MOTION OF IONS IN 3-DIMENSIONAL QUADRUPOLE ELECTRIC-FIELDS
    BONNER, RF
    MARCH, RE
    DURUP, J
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1976, 22 (1-2): : 17 - 34
  • [9] Determination of 236U/238U isotope ratio in contaminated environmental samples using different ICP-MS instruments
    Boulyga, SF
    Matusevich, JL
    Mironov, VP
    Kudrjashov, VP
    Halicz, L
    Segal, I
    McLean, JA
    Montaser, A
    Becker, JS
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (08) : 958 - 964
  • [10] Optimisation and application of ICP-MS and alpha-spectrometry for determination of isotopic ratios of depleted uranium and plutonium in samples collected in Kosovo
    Boulyga, SF
    Testa, C
    Desideri, D
    Becker, JS
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (11) : 1283 - 1289