Determination of cobalt in seawater by catalytic adsorptive cathodic stripping voltammetry

被引:104
作者
Vega, M [1 ]
van den Berg, CMG [1 ]
机构
[1] UNIV LIVERPOOL, OCEANOG LABS, LIVERPOOL L69 3BX, MERSEYSIDE, ENGLAND
关键词
D O I
10.1021/ac960214s
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new procedure for the direct determination of picomolar levels of cobalt in seawater is presented. Cathodic stripping voltammetry is preceded by adsorptive accumulation of the cobalt-nioxime (cyclohexane-1,2-dione dioxime) complex from seawater containing 6 mu M nioxime and 80 mM ammonia at pH 9.1, onto a hanging mercury drop electrode, followed by reduction of the adsorbed species. The reduction current is catalytically enhanced by the presence of 0.5 M nitrite. Optimized conditions for cobalt include a 30 s adsorption period at -0.7 V and a voltammetric scan using differential pulse modulation, According to the proposed reaction mechanism, dissolved Co(Il) is oxidized to Co(III) upon addition of nioxime and high concentrations of ammonia and nitrite; a mixed Co(III)-ammonia-nitrite complex is adsorbed on the electrode surface; the Co(III) is reduced to Co(II) (complexed by nioxime) during the voltammetric scan, followed by its chemical reoxidation by the nitrite, initiating a catalytically enhanced current. A detection limit of 3 pM cobalt (at an adsorption period of 60 s) enables the detection of this metal in uncontaminated seawater using a very short adsorption time, UV digestion of seawater is essential, as part of the cobalt may occur strongly complexed by organic matter and rendered nonlabile. The method was applied successfully to the determination of the distribution of cobalt in the water column of the Mediterranean.
引用
收藏
页码:874 / 881
页数:8
相关论文
共 43 条
[1]  
ACHTERBERG EP, IN PRESS DEEP SEA RE
[2]   THE SURFACE-CHEMISTRY OF SEDIMENTS FROM THE PANAMA BASIN - THE INFLUENCE OF MN OXIDES ON METAL ADSORPTION [J].
BALISTRIERI, LS ;
MURRAY, JW .
GEOCHIMICA ET COSMOCHIMICA ACTA, 1986, 50 (10) :2235-2243
[3]   EXPLOITATION OF THE NITRITE CATALYTIC EFFECT TO ENHANCE THE SENSITIVITY AND SELECTIVITY OF THE ADSORPTIVE STRIPPING VOLTAMMETRIC METHOD FOR THE DETERMINATION OF COBALT WITH DIMETHYLGLYOXIME [J].
BOBROWSKI, A ;
BOND, AM .
ELECTROANALYSIS, 1992, 4 (10) :975-979
[4]   DETERMINATION OF COBALT BY ADSORPTIVE STRIPPING VOLTAMMETRY USING COBALT(II)-NIOXIME-NITRITE CATALYTIC-SYSTEM [J].
BOBROWSKI, A .
ANALYTICAL LETTERS, 1990, 23 (08) :1487-1503
[6]   THE DETERMINATION OF THE CHROMIUM SPECIATION IN SEA-WATER USING CATALYTIC CATHODIC STRIPPING VOLTAMMETRY [J].
BOUSSEMART, M ;
VAN DEN BERG, CMG ;
GHADDAF, M .
ANALYTICA CHIMICA ACTA, 1992, 262 (01) :103-115
[7]  
Bowen H.J.M., 1966, TRACE ELEMENTS BIOCH
[8]   SAMPLING AND ANALYTICAL METHODS FOR THE DETERMINATION OF COPPER, CADMIUM, ZINC, AND NICKEL AT THE NANOGRAM PER LITER LEVEL IN SEA-WATER [J].
BRULAND, KW ;
FRANKS, RP ;
KNAUER, GA ;
MARTIN, JH .
ANALYTICA CHIMICA ACTA, 1979, 105 (01) :233-245
[9]  
BRULAND KW, 1983, CHEM OCEANOGR, V8, pCH45
[10]  
CARVALHO FP, 1986, HEALTH PHYS, V53, P73