Recognition of electrocardiograph left arm left leg lead reversal

被引:42
作者
Abdollah, H [1 ]
Milliken, JA [1 ]
机构
[1] QUEENS UNIV,DIV CARDIOL,HOP HOTEL DIEU,KINGSTON,ON K7L 5G2,CANADA
关键词
D O I
10.1016/S0002-9149(97)00656-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The electrocardiographic error of left arm/left leg lead reversal is difficult to identify. PI amplitude greater than PII as a terminal positive component to PIII may diagnose 90% of such errors.
引用
收藏
页码:1247 / &
页数:4
相关论文
共 5 条
  • [1] ELECTROCARDIOGRAPHIC PATTERNS RESULTING FROM IMPROPER CONNECTIONS OF THE RIGHT LEG (GROUND) CABLE
    CASTELLANOS, A
    SAOUDI, NC
    SCHWARTZ, A
    SODIPALLARES, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (03): : 364 - 368
  • [2] ECG DIAGNOSTIC ERRORS DUE TO IMPROPER CONNECTION OF THE RIGHT ARM AND LEG CABLES
    GUIJARROMORALES, A
    GILEXTREMERA, B
    MALDONADOMARTIN, A
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 1991, 30 (02) : 233 - 235
  • [3] RECOGNITION OF ELECTROCARDIOGRAPHIC ELECTRODE MISPLACEMENTS INVOLVING THE GROUND (RIGHT LEG) ELECTRODE
    HAISTY, WK
    PAHLM, O
    EDENBRANDT, L
    NEWMAN, K
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (16) : 1490 - 1495
  • [4] ARTIFICIAL NEURAL NETWORKS FOR RECOGNITION OF ELECTROCARDIOGRAPHIC LEAD REVERSAL
    HEDEN, B
    OHLSSON, M
    EDENBRANDT, L
    RITTNER, R
    PAHLM, O
    PETERSON, C
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1995, 75 (14) : 929 - 933
  • [5] Detection of frequently overlooked electrocardiographic lead reversals using artificial neural networks
    Heden, B
    Ohlsson, M
    Holst, H
    Mjoman, M
    Rittner, R
    Pahlm, O
    Peterson, C
    Edenbrandt, L
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (05) : 600 - &