Evaluation of a commercial three-dimensional electron beam treatment planning system

被引:19
作者
Ding, GX
Cygler, JE
Zhang, GG
Yu, MK
机构
[1] British Columbia Canc Agcy, Fraser Valley Canc Ctr, Surrey, BC V3V 1Z2, Canada
[2] Ottawa Reg Canc Ctr, Dept Phys Med, Ottawa, ON K1Y 4K7, Canada
[3] Natl Res Council Canada, Ottawa, ON, Canada
关键词
electron beam; treatment planning; inhomogeneous phantom;
D O I
10.1118/1.598795
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We evaluated a commercial three-dimensional (3D) electron beam treatment planning system (CADPLAN V.2.7.9) using both experimentally measured and Monte Carlo calculated dose distributions to compare with those predicted by CADPLAN calculations. Tests were carried out at various field sizes and electron beam energies from 6 to 20 MeV. For a homogeneous water phantom the agreement between measured and CADPLAN calculated dose distributions is very good except at the phantom surface. CADPLAN is able to predict hot and cold spots caused by a simple 3D inhomogeneity but unable to predict dose distributions for a more complex geometry where CADPLAN underestimates dose changes caused by inhomogeneity. We discussed possible causes for the inaccuracy in the CADPLAN dose calculations. In addition, we have tested CADPLAN treatment monitor unit and electron cut-out factor calculations and found that CADPLAN predictions generally agree with manual calculations. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00412-5].
引用
收藏
页码:2571 / 2580
页数:10
相关论文
共 16 条
[1]  
[Anonymous], 1984, 35 ICRU
[2]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[3]  
Ding G, 1995, PIRS, V439
[4]  
HOGSTROM KR, TELETHERAPY PRESENT, P137
[5]  
HOGSTROM KR, ADV RAD ONCOLOGY PHY, P900
[6]  
HYODYNMAA S, 1991, TECHNICAL RES CTR FI, V74
[7]  
ICRU, 1984, 37 ICRU
[8]  
Lax I, 1983, Acta Radiol Suppl, V364, P49
[9]  
LAX I, 1986, THESIS STOCKHOLM U S
[10]   BEAM - A MONTE-CARLO CODE TO SIMULATE RADIOTHERAPY TREATMENT UNITS [J].
ROGERS, DWO ;
FADDEGON, BA ;
DING, GX ;
MA, CM ;
WE, J ;
MACKIE, TR .
MEDICAL PHYSICS, 1995, 22 (05) :503-524