Rapid HbA(1c) testing in a community setting

被引:27
作者
Carter, JS
Houston, CA
Gilliland, SS
Perez, GE
Owen, CL
Pathak, DR
Little, RR
机构
[1] VET ADM, ALBUQUERQUE, NM USA
[2] MICHIGAN STATE UNIV, E LANSING, MI 48824 USA
[3] ALL ST SENIOR HLTH CTR, HALTOM CITY, TX USA
[4] UNIV MISSOURI, DEPT PATHOL & ANAT SCI, COLUMBIA, MO USA
[5] UNIV MISSOURI, DEPT CHILD HLTH, COLUMBIA, MO 65201 USA
关键词
D O I
10.2337/diacare.19.7.764
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE - To determine whether the DCA 2000 analyzer provides valid and reliable HbA(1c) results when used under field conditions and operated by nonmedical personnel. This study was part of a community diabetes education program, the Native American Diabetes Project, in which HbA(1c) was measured as an indicator of average glycemic control. RESEARCH DESIGN AND METHODS - Two study samples were taken, the first in the spring of 1994 and the second in the spring of 1995. Seven community members in 1994 and six new community members in 1995 were trained over 2 days, using standard protocol, to operate the DCA 2000 HbA(1c) analyzer and to collect two capillary blood samples from participants in the Native American Diabetes Project. Duplicate DCA 2000 HbA(1c) measurements performed by the community workers were compared with measurements from a highperformance liquid chromatography (HPLC) system. Validity and reliability measures were calculated. RESULTS - Of the participants, 43 were studied in 1994 and 14 in 1995. Comparison of the mean DCA 2000 results with those of HPLC showed high validity, with the absolute relative difference between the mean DCA 2000 and the external reference of HPLC (\ mean DCA 2000 - HPLC \ /HPLC) as 4.0 and 2.0% for 1994 and 1995, respectively. The Pearson correlation coefficients (r) between these two measures were 0.968 and 0.996 for 1994 and 1995, respectively. While the 1994 data appeared to have less validity for values >10%, they included only one value with a 60-min warm-up of the DCA analyzer. The 1995 data, all collected after a 60-min warm-up, had good correlation throughout the range of values. The within-run reliability was excellent, with an intraclass correlation coefficient of reliability of 0.959 and 0.975 for paired samples, for 1994 and 1995 respectively. The mean coefficient of variation for these paired measures was 3.0% in 1994 and 2.8% in 1995. Both validity and reliability were improved by changing the warm-up period of the DCA 2000 analyzer from 5 to 60 min. All correlation coefficients were statistically significant (P < 0.0001). CONCLUSIONS - The DCA 2000 gave valid and reliable HbA(1c) results when operated in a community setting by nonmedical personnel. Extending the warm-up period of the device to 60 min slightly improved the validity and reliability of the test.
引用
收藏
页码:764 / 767
页数:4
相关论文
共 9 条
  • [1] EASTMAN RC, 1994, DIABETES REV, V2, P263
  • [2] Fleiss J., 1986, Reliability of measurement: the design and analysis of clinical experiments
  • [3] Goldstein DE, 1986, METHODS DIABETES RES, V2, P475
  • [4] A MULTISITE PHYSICIANS OFFICE LABORATORY EVALUATION OF AN IMMUNOLOGICAL METHOD FOR THE MEASUREMENT OF HBA1C
    GUTHRIE, R
    HELLMAN, R
    KILO, C
    HIAR, CE
    CROWLEY, LE
    CHILDS, B
    FISHER, R
    PINSON, MB
    SUTTNER, A
    VITTORI, C
    [J]. DIABETES CARE, 1992, 15 (11) : 1494 - 1498
  • [5] Hulley SB, 2007, Designing clinical research: an epidemiologic approach, VFourth
  • [6] LITTLE RR, 1992, CLIN CHEM, V38, P2472
  • [7] IMMEDIATE HBA1C RESULTS - PERFORMANCE OF NEW HBA1C SYSTEM IN PEDIATRIC OUTPATIENT POPULATION
    MARRERO, DG
    VANDAGRIFF, JL
    GIBSON, R
    FINEBERG, SE
    FINEBERG, SE
    FINEBERG, NS
    HIAR, CE
    CROWLEY, LE
    [J]. DIABETES CARE, 1992, 15 (08) : 1045 - 1049
  • [8] THE EFFECT OF INTENSIVE TREATMENT OF DIABETES ON THE DEVELOPMENT AND PROGRESSION OF LONG-TERM COMPLICATIONS IN INSULIN-DEPENDENT DIABETES-MELLITUS
    SHAMOON, H
    DUFFY, H
    FLEISCHER, N
    ENGEL, S
    SAENGER, P
    STRELZYN, M
    LITWAK, M
    WYLIEROSETT, J
    FARKASH, A
    GEIGER, D
    ENGEL, H
    FLEISCHMAN, J
    POMPI, D
    GINSBERG, N
    GLOVER, M
    BRISMAN, M
    WALKER, E
    THOMASHUNIS, A
    GONZALEZ, J
    GENUTH, S
    BROWN, E
    DAHMS, W
    PUGSLEY, P
    MAYER, L
    KERR, D
    LANDAU, B
    SINGERMAN, L
    RICE, T
    NOVAK, M
    SMITHBREWER, S
    MCCONNELL, J
    DROTAR, D
    WOODS, D
    KATIRGI, B
    LITVENE, M
    BROWN, C
    LUSK, M
    CAMPBELL, R
    LACKAYE, M
    RICHARDSON, M
    LEVY, B
    CHANG, S
    HEINHEINEMANN, M
    BARRON, S
    ASTOR, L
    LEBECK, D
    BRILLON, D
    DIAMOND, B
    VASILASDWOSKIN, A
    LAURENZI, B
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1993, 329 (14) : 977 - 986
  • [9] 1995, DIABETES CARE, V18, P8