Comparisons of platinum, gold, palladium and glassy carbon as electrode materials in the design of biosensors for glutamate

被引:109
作者
O'Neill, RD
Chang, SC
Lowry, JP
McNeil, CJ
机构
[1] Univ Coll Dublin, Dept Chem, Dublin 4, Ireland
[2] Univ Newcastle Upon Tyne, Inst Nanoscale Sci & Technol, Sch Med, Sch Clin & Lab Sci, Newcastle Upon Tyne, Tyne & Wear, England
[3] Natl Univ Ireland, Dept Chem, Maynooth, Kildare, Ireland
关键词
poly( 1,2-diaminobenzene); poly(phenylenediamine); glutamate oxidase; horseradish peroxidase; Os2+-polyvinylpyridine; ascorbic acid; amperometric detection; cyclic voltammetry; hydrogen peroxide;
D O I
10.1016/j.bios.2003.12.004
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Four electrode materials: Pt, An, Pd and glassy carbon (GC), were studied to investigate their suitability as substrates in the development of two different classes of glutamate biosensor. Glutamate oxidase cross-linked onto poly(o-phenylenediamine) was chosen as the type I biosensor (PPD/GluOx), incorporating PPD as the permselective element to detect H2O2 directly on the electrode surface at relatively high applied potentials. GluOx and horseradish peroxidase/redox polymer modified electrodes (Os2+ PVP/HRP/GluOx) that relied on enzyme-catalysed H2O2, detection at lower applied potentials were used as type 2 biosensors. The voltammetric and amperometric responses to the enzyme signal transduction molecule, H2O2 and the archetypal interference species in biological applications, ascorbic acid, were determined on the bare and PPD/GluOx-modified surfaces. The amperometric responses of these electrodes were stable over several days of continuous recording in phosphate buffered saline (pH 7.4). The sensitivity of the type I biosensors to H2O2 and glutamate showed parallel trends with low limits of detection and good linearity at low concentrations: Pt > An similar to Pd much greater than GC. Type 2 biosensors out-performed the type I design for all electrode substrates, except Pt. However, the presence of the permselective PPD membrane in the type I biosensors, not feasible in the type 2 design, suggests that Pt/PPD/GluOx might have the best all-round characteristics for glutamate detection in biological media containing interference species such as ascorbic acid. Other points affecting a final choice of substrate should include factors such as mass production issues. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:1521 / 1528
页数:8
相关论文
共 55 条
[1]  
[Anonymous], 1997, BIOSENSORS BODY CONT
[2]   An enzyme switch employing direct electrochemical communication between horseradish peroxidase and a poly(aniline) film [J].
Bartlett, PN ;
Birkin, PR ;
Wang, JH ;
Palmisano, F ;
De Benedetto, G .
ANALYTICAL CHEMISTRY, 1998, 70 (17) :3685-3694
[3]   STUDIES ON ENZYME ELECTRODES WITH FERROCENE AND CARBON PASTE BOUND WITH CELLULOSE TRIACETATE [J].
BEH, SK ;
MOODY, GJ ;
THOMAS, JDR .
ANALYST, 1991, 116 (05) :459-462
[4]   FAST ELECTROCHEMICAL REACTIONS DURING THE ANODIZATION OF PLATINUM-ELECTRODES IN AQUEOUS-SOLUTIONS - DETECTION OF A REVERSIBLE CHARGE STORAGE PROCESS AT PLATINUM-OXIDE-COATED ELECTRODES [J].
BOLZAN, AE ;
ARVIA, AJ .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1992, 341 (1-2) :93-109
[5]   PREMONOLAYER OXIDATION AND ITS ROLE IN ELECTROCATALYSIS [J].
BURKE, LD .
ELECTROCHIMICA ACTA, 1994, 39 (11-12) :1841-1848
[6]   Anomalous oxidation reactions at noble metal surfaces at low potentials: With particular reference to palladium [J].
Burke, LD ;
Buckley, DT .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1996, 143 (03) :845-854
[7]   Improved ceramic-based multisite microelectrode for rapid measurements of L-glutamate in the CNS [J].
Burmeister, JJ ;
Pomerleau, F ;
Palmer, M ;
Day, BK ;
Huettl, P ;
Gerhardt, GA .
JOURNAL OF NEUROSCIENCE METHODS, 2002, 119 (02) :163-171
[8]   PERMEATION OF SOLUTES THROUGH AN ELECTROPOLYMERIZED ULTRATHIN POLY-O-PHENYLENEDIAMINE FILM USED AS AN ENZYME-ENTRAPPING MEMBRANE [J].
CENTONZE, D ;
MALITESTA, C ;
PALMISANO, F ;
ZAMBONIN, PG .
ELECTROANALYSIS, 1994, 6 (5-6) :423-429
[9]   Determination of glutamate pyruvate transaminase activity in clinical specimens using a biosensor composed of immobilized L-glutamate oxidase in a photo-crosslinkable polymer membrane on a palladium-deposited screen-printed carbon electrode [J].
Chang, KS ;
Hsu, WL ;
Chen, HY ;
Chang, CK ;
Chen, CY .
ANALYTICA CHIMICA ACTA, 2003, 481 (02) :199-208
[10]   ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY [J].
CLARK, LC ;
LYONS, C .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) :29-&