Membrane-assisted solvent extraction of triazines and other semi-volatile contaminants directly coupled to large-volume injection-gas chromatography-mass spectrometric detection

被引:68
作者
Hauser, B
Popp, P
Kleine-Benne, E
机构
[1] UFZ Helmholtz Ctr Environm Res, Dept Analyt Chem, D-04318 Leipzig, Germany
[2] Gerstel GmbH & Co KG, D-45473 Mulheim, Germany
关键词
membrane-assisted solvent extraction; extraction methods; large-volume injection; instrumentation; triazines;
D O I
10.1016/S0021-9673(02)00135-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple device was developed for in-vial liquid-liquid extraction using a polymer membrane (nonporous polypropylene) to separate an aqueous sample from an organic extractant. The membrane consisted of tubing with an internal diameter of 6 mm and a wall thickness of 0.05 mm, which was heat-sealed at the lower end and filled with 500 mul hexane. This membrane bag was incorporated into a conventional 20 ml headspace vial suitable for a multi-purpose sampler (NIPS 2, Gerstel, Mulheim, Germany) directly interfaced to a gas chromatograph with a mass-selective detector. The sampler enabled the extraction vial to be mixed at a defined temperature with subsequent large-volume injection of the organic extract taken from the membrane bag. The method was evaluated using several triazines, 2,4-dichloroaniline, alpha-hexachlorocyclohexane and phenanthrene as model compounds. Extraction parameters such as temperature, agitation speed, and extraction time were optimised. Recoveries of 60-90% were achieved after 30 min extraction. By increasing the injection volume to 100 mul, detection limits of 1-10 ng/l were determined. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:27 / 36
页数:10
相关论文
共 27 条
[1]  
Halvorsen TG, 2001, J SEP SCI, V24, P615, DOI 10.1002/1615-9314(20010801)24:7<615::AID-JSSC615>3.0.CO
[2]  
2-Z
[3]  
Hauser B, 2001, J SEP SCI, V24, P551, DOI [10.1002/1615-9314(20010801)24:7<551::AID-JSSC551>3.0.CO
[4]  
2-2, 10.1002/1615-9314(20010801)24:7&lt
[5]  
551::AID-JSSC551&gt
[6]  
3.0.CO
[7]  
2-]
[8]   Liquid membrane extraction in analytical sample preparation I.: Principles [J].
Jönsson, JÅ ;
Mathiasson, L .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1999, 18 (05) :318-325
[9]   Liquid membrane extraction in analytical sample preparation II.: Applications [J].
Jönsson, JÅ ;
Mathiasson, L .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1999, 18 (05) :325-334
[10]  
Laespada EF, 1998, J CHROMATOGR A, V823, P537