Measurement variation among 12 electronic home blood pressure monitors

被引:42
作者
Yarows, SA [1 ]
Brook, RD [1 ]
机构
[1] Univ Michigan Hlth Syst, Div Hypertens Internal Med, Ann Arbor, MI USA
关键词
blood pressure; self-care; home blood pressure devices; validation; standards;
D O I
10.1016/S0895-7061(99)00182-X
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Although there are AAMI and BHS standards for accuracy of electronic home electronic blood pressure monitors (HBPM), patient composition differences and differences in manufacturer's algorithm for calculation of the systolic and diastolic measurement may result in measurement differences between monitors. The aim of this study was the measurement of differences among HBPM. Paired comparisons were performed between the Omron 712c electronic home monitor and each of 12 other HBPM (Sunbeam 7654, Sunbeam 7623, Omron 711, Omron 432c, A&D-UA767, Lumiscope 1085M, Omron 725CIC, Assure A30, Lumiscope 1083N, Omron 815, Omron 605, and Assure BDW20), in addition to comparison to the auscultatory method by trained observers. Measurements were made in normotensive subjects in an ambulatory setting. The main outcome measures were systolic and diastolic blood pressure measurements. All of the HBPM, except for the Sunbeam 7654 and the Assure A30/BD-W20 (wrist) models, demonstrated small differences of <4/4 mm Hg for systolic/diastolic measurements with pulse measurement differences of <3 beats/min. These differences were less than the differences previously reported fox office BP auscultation of 6/5-10 mm Hg for systolic/diastolic measurements. The Omron 712c, passing previous AAMI and BHS standards, measured the systolic reading within 2 mm Hg of auscultatory mercury or aneroid measurement and under-measured the diastolic by 6-9 mm Hg, Differences in the patient composition studied could account for the difference. The wrist and finger manometers performed clinically similar to the Omron 712c, except for the Assure BD-W20, which over-measured the diastolic by 7 mm Hg. It is concluded that the small differences among the various HBPM, which are less than those in clinical office auscultation, should encourage greater use of electronic manometers in the office and at home. Am J Hypertens 2000;13:276-282 (C) 2000 American Journal of Hypertension, Ltd.
引用
收藏
页码:276 / 282
页数:7
相关论文
共 39 条
  • [1] *ASS ADV MED INSTR, 1992, EL AUT SPHYGM, P1
  • [2] BOROTOLOTTO LA, 1999, BLOOD PRESS MONIT, V4, P21
  • [3] Cordoba R, 1997, Aten Primaria, V20, P247
  • [4] Eckert S, 1996, Z KARDIOL, V85, P109
  • [5] EVANS CE, 1989, J HYPERTENS, V7, P133
  • [6] FOSTER C, 1994, J HUM HYPERTENS, V8, P661
  • [7] Effects of intensive blood-pressure lowering and low-dose aspirin in patients with hypertension:: principal results of the hypertension optimal treatment (HOT) randomised trial
    Hansson, L
    Zanchetti, A
    Carruthers, SG
    Dahlöf, B
    Elmfeldt, D
    Julius, S
    Ménard, J
    Rahn, KH
    Wedel, H
    Westerling, S
    [J]. LANCET, 1998, 351 (9118) : 1755 - 1762
  • [8] HARRISON DW, 1987, MED INSTRUM, V21, P323
  • [9] CLINICAL-EVALUATION OF SEMIAUTOMATIC AND AUTOMATIC DEVICES FOR HOME BLOOD-PRESSURE MEASUREMENT - COMPARISON BETWEEN CUFF-OSCILLOMETRIC AND MICROPHONE METHODS
    IMAI, Y
    ABE, K
    SASAKI, S
    MINAMI, N
    MUNAKATA, M
    SAKUMA, H
    HASHIMOTO, J
    SEKINO, H
    IMAI, K
    YOSHINAGA, K
    [J]. JOURNAL OF HYPERTENSION, 1989, 7 (12) : 983 - 990
  • [10] IYRIBOZ Y, 1990, J FAM PRACTICE, V31, P376