Exploring the limits of RF shimming for high-field MRI of the human head

被引:193
作者
Mao, Weihua [1 ]
Smith, Michael B. [1 ]
Collins, Christopher M. [1 ]
机构
[1] Penn State Univ, Coll Med, Dept Radiol, Ctr NMR Res, Hershey, PA 17033 USA
关键词
MRI; high field; array; RF shimming; simulations;
D O I
10.1002/mrm.21013
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Several methods have been proposed for overcoming the effects of radiofrequency (RF) magnetic field inhomogeneity in high-field MRI. Some of these methods rely at least in part on the ability to independently control magnitude and phase of different drives in either one multielement RF coil or in different RF coils in a transmit array. The adjustment of these drive magnitudes and phases alone to create uniform RF magnetic (B-1) fields has been called RF shimming, and has certain limits at every frequency as dictated by possible solutions to Maxwell's equations. Here we use numerical calculations to explore the limits of RF shimming in the human head. We found that a 16-element array can effectively shim a single slice at frequencies up to 600 MHz and the whole brain at up to 300 MHz, while an 80-element array can shim the whole brain at up to 600 MHz.
引用
收藏
页码:918 / 922
页数:5
相关论文
共 18 条
[1]   Transmit and receive transmission line arrays for 7 tesla parallel imaging [J].
Adriany, G ;
Van de Moortele, PF ;
Wiesinger, F ;
Moeller, S ;
Strupp, JP ;
Andersen, P ;
Snyder, C ;
Zhang, XL ;
Chen, W ;
Pruessmann, KP ;
Boesiger, P ;
Vaughan, T ;
Ugurbil, K .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (02) :434-445
[2]  
[Anonymous], 1997, Compilation of the Dielectric Properties of Body Tissues at RF and Microwave Frequencies
[3]   Combination of optimized transmit arrays and some receive array reconstruction methods can yield homogeneous images at very high frequencies [J].
Collins, CM ;
Liu, W ;
Swift, BJ ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (06) :1327-1332
[4]   Signal-to-noise ratio and absorbed power as functions of main magnetic field strength, and definition of "90°" RF pulse for the head in the birdcage coil [J].
Collins, CM ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (04) :684-691
[5]  
COLLINS CM, 2004, P 12 ANN M ISMRM KYO, P1566
[6]   The NMR multi-transmit phased array: a Cartesian feedback approach [J].
Hoult, DI ;
Kolansky, G ;
Kripiakevich, D ;
King, SB .
JOURNAL OF MAGNETIC RESONANCE, 2004, 171 (01) :64-70
[7]  
Hoult DI, 2000, J MAGN RESON IMAGING, V12, P46, DOI 10.1002/1522-2586(200007)12:1<46::AID-JMRI6>3.0.CO
[8]  
2-D
[9]  
IBRAHIM T, 2005, P 13 ANN M ISMRM MIA, P2749
[10]  
KUPRAD KN, 2005, P 13 ANN M ISMRM MIA, P16